Can a 10 MHz OCXO reuse yesterday's calibration after power was off overnight?

A measurement instrument still outputs 10 MHz after being switched off overnight. Production wants to reuse yesterday's calibration, while the site wants to start measuring within minutes of power-on. Evaluate those requirements separately when selecting the OCXO. First output, entry into the warm-up band and return to the previous frequency can all occur at different times.

Find the two frequency references in the datasheet

Rakon's STP3558LF specification covers a 10 MHz, 3.3 V ROX2522S3 without external voltage tuning. Section 6 defines warm-up as reaching ±100 ppb of the frequency after 24 hours of continuous operation. At 25 °C, that configuration has a maximum warm-up time of five minutes. Section 8 separately specifies frequency retrace at ±50 ppb after 24 hours powered on, 24 hours powered off and one hour powered on again. Retrace measures how closely frequency returns to its former value after the cycle. Each specification has its own measurement time and reference.Rakon STP3558LF datasheet, Sections 6 and 8

At 10 MHz, 100 ppb is 1 Hz and 50 ppb is 0.5 Hz. If equipment needs to be within ±0.5 Hz of yesterday's frequency five minutes after power-on, the five-minute warm-up specification and one-hour retrace specification neither individually establish that combined requirement. Obtain a guarantee for the specific time or include it in system testing. An instrument that permits precision measurements only after one hour has a different evidence requirement.

Distinguish return to the previous frequency from reaching nominal

Suppose the pre-shutdown reading is 10,000,000.80 Hz. After the agreed outage and at the specified measurement time after power returns, the reading is 10,000,001.10 Hz. The increase is +0.30 Hz, or +30 ppb relative to nominal. The second reading itself is +1.10 Hz above nominal, or +110 ppb. Record the first difference for retrace. If the instrument must produce an accurate absolute reference, check the second statistic and the reference timebase uncertainty too. Repeatable return to the old frequency does not establish that the old frequency was correct.

Switch off the device under test independently. Keep the reference instrument running and retain its lock status. Hold output load, supply settings and sampling window fixed, save the last continuous readings before power-off, and log again from the first output after restoration. Record separate times for power loss, power restoration and repeat measurement so one hour from logging is not mistaken for one hour from restoration. If the system has automatic calibration, record whether and when it starts correcting, since it can affect the observed curve.

An overnight outage and a month in storage require different evidence

Before selecting parts, ask how long equipment can stay powered off and which operation first requires accurate frequency after restart. Verify normal overnight shutdown for an instrument that uses that daily schedule. Separately measure restart after long outages for spares that may spend weeks in storage. A datasheet value after one day off cannot establish behavior after several weeks. Record storage temperature separately. For multiple samples, keep individual identifiers and starting baselines to separate unit-to-unit differences from cycle-to-cycle variation.

Write an executable purchasing requirement. For example, at specified supply and environmental conditions, power the unit for a specified time and save the baseline, remove power for a specified time, then read the frequency at the specified minute after power returns and compare it with the baseline. Include the allowable difference, averaging window and method for retesting borderline results. Then engineering can evaluate whether a replacement labeled 10 MHz permits the same release time after startup.