A frequency counter reads several hertz lower after supply falls from 3.45 to 3.15 V. The change may include supply pushing, receiver level sensitivity, temperature, or the measurement path. Decide whether the requirement concerns settled operation at different voltages or transient excursions after a step. The windows differ. If there is a separate tuning pin, fix and verify its actual voltage during supply sweeps.
Two points give an average slope over that interval
Suppose the settled readings are 24,999,998 Hz at 3.15 V and 25,000,002 Hz at 3.45 V. The difference is +4 Hz across 0.30 V, about +13.33 Hz/V or +0.533 ppm/V at 25 MHz. This describes the sample's average over that interval under the test conditions. Add intermediate voltages and returns to nominal to detect a flatter center or steeper edge before extrapolating to small changes.
If a local slope of 13.33 Hz/V is independently verified near 3.30 V, a quasi-static ±40 mV variation predicts about ±0.53 Hz or ±0.021 ppm. Fast ripple requires a separate dynamic measurement. Regulators, bias, and compensation can alter the response with modulation frequency. Verify the frequency effect independently of output amplitude and threshold changes.
Measure the voltage at the package
Use low-noise supply within rated conditions. Measure across device supply and ground rather than assuming the display represents pin voltage. Cables, protection devices, and regulators can drop more voltage at different loads. Record current and output level with frequency. If output falls at low voltage, ensure the counter still has threshold margin, and cross-check by another method if needed. Keep device loading unchanged during threshold diagnostics.
Try nominal, low, nominal, high, nominal. Use documented waiting and identical gate lengths and log temperature and frequency through each step. An immediate 2 Hz jump followed by another 2 Hz drift over minutes deserves separate examination. The latter could be thermal or related to internal compensation or power. If case or ambient temperature changes, repeat with equivalent thermal conditions or establish a suitable shorter observation window. Reversing order can reveal warm-up or aging during the sweep.
Record the supply-induced and absolute results
Suppose the supply specification allows ±1 Hz relative to the nominal supply reading. At 25,000,000 Hz nominal, the two example endpoints are each 2 Hz away. Evaluate that requirement using the agreed rule. For a system absolute frequency limit, retain actual readings alongside temperature, loading, warm-up, and uncertainty. For a VCXO, identify which measurements vary supply and which vary the tuning control. Keep all returns to nominal to see whether thermal and drift effects contaminated the comparisons.
Document thermometer locations. An air sensor and a device-adjacent sensor can reveal continued board heating after chamber air stabilizes. If attaching to a tiny package would alter its thermal state, use a reproducible nearby point and identify the difference. Verify the setup at room temperature before repeating at corners. Keep loads, output mode, and cycling sequence consistent. Compare frequency and temperature settling when self-heating is appreciable. A supply-induced power change can create a different long-term thermal equilibrium relevant to service.
Check resolution before interpreting slope changes. A 0.1 Hz step with 0.4 Hz scatter at a fixed voltage does not yield a precise derivative. Use several nearby voltages and independent repeats with fitted residuals. Record instrument reference or counting-range transitions so an instrument discontinuity is not fitted as a device characteristic.