Why changing coax can shift a 80 MHz oscillator power reading by 0.3 dB

A sine-output oscillator drives an RF power sensor. Two coaxial cables give different readings even after insertion-loss correction. Check the frequency, sensor range, and sensor calibration factor. Then examine mismatch. Nominal 50 Ω ports still reflect at the operating frequency. A different cable length changes round-trip phase and the way the returned signal combines at the sensor. An insertion-loss correction alone leaves that term.

Estimate the possible swing

For a single-frequency example, take 14 dB source return loss and 20 dB sensor return loss. The reflection magnitude is 10^(−return loss/20), giving about 0.200 and 0.100. Approximating the cable as lossless gives a round-trip reflection product magnitude of 0.020. The reflected-wave power factor is 1/|1−ΓsourceΓload|² with the load reflection referred to the source plane. Its phase-dependent bounds are about −0.172 and +0.175 dB, from −20 log₁₀(1+0.020) and −20 log₁₀(1−0.020). Two cables near opposite phase extremes could differ by about 0.35 dB from this term alone. Sensor absorption, calibration factor, insertion loss, and any more complex source response are separate terms.

The example bounds a plausible change. It does not predict its sign for a given cable. Connectors and actual electrical length matter even for cables marked with the same length. If power trends during a temperature sweep, investigate the cable, sensor, and reference source alongside the oscillator itself.

Check the bench

Keep supply, frequency, and connectors fixed. Alternate between the cables twice and check how consistently the difference returns after remating. Respect the sensor's input limits. Characterize the passive cable, pad, and permissible sensor reflection with a suitable calibrated network analyzer at the required reference planes and power levels. An operating oscillator generates its own signal. Measuring its source reflection requires a method suited to a live source and appropriate instrument protection. When such a setup is unavailable, use the specified source match at that frequency or a justified bound.

Where oscillator loading permits, repeat the comparison with a well-matched calibrated pad before the sensor. An ideal 10 dB pad gives reflected waves 20 dB of round-trip attenuation and reduces the reflection coefficient seen at the source by 10^(−20/20), or about 0.100. The example product falls from about 0.020 to 0.0020, yielding about 0.035 dB from one phase extreme to the other if the pad's own reflection is small. The sensor also loses about 10 dB of forward signal. Verify sensitivity, linear range, pad calibration, and dynamic range. A smaller cable effect with the pad supports a mismatch contribution. Since the pad changes oscillator loading, check frequency and waveform in parallel if necessary.

Report the plane and the power definition

Suppose the acceptance range is +4 to +10 dBm for steady sine power into a specified 50 Ω load at the device port. If the corrected reading is +4.10 dBm and its interval considering mismatch, cable, and sensor terms crosses +4 dBm, apply the agreed decision rule. Log insertion losses, pad calibration, sensor settings, return-loss sources, and reference planes. A broadband sensor can include harmonic power. If the specification covers fundamental power, use a frequency-selective method or quantify the harmonic contribution. A repeat test can then use the same quantity.

Separate connector repeatability from cable dependence. Remate one cable several times using a consistent procedure, then compare the two cables and repeat both checks with the pad. If remating a connector alone moves the result by several tenths of a decibel, check cleanliness, wear, and torque before interpreting it as a length effect. If each cable repeats reliably and the difference between cables returns, investigate reflection and phase. Run the same load configuration with a check source to see whether cable dependence persists.

Allow for short-term drift. Two ten-minute measurements may include warm-up movement of the source or meter. A short-long-short sequence reveals part of any change that does not reverse. Keep the frequency calibration factor and reference plane consistent when repeating with a pad so the reduction in setup sensitivity can be assessed.