How to budget uncertainty for a 0.7 Hz difference between two 10 MHz oscillators

A counter reads 10,000,001.2 Hz for A and 10,000,001.9 Hz for B. Its timebase uncertainty may look comparable to the 0.7 Hz difference. First define the quantity: absolute frequency, simultaneous difference, or difference from successive readings. A common error cancels in the difference only to the extent it affects both measurements alike.

A paired difference can have less uncertainty than either reading

Suppose each reading has 0.50 Hz standard uncertainty. For d = B−A, the variance is uA² + uB² − 2cov(A,B). With a justified error correlation of 0.80, covariance is 0.80 × 0.50 × 0.50 = 0.20 Hz². Then u(d) = √(0.25+0.25−0.40), or about 0.316 Hz. For independent contributions with the same individual uncertainties, it is about 0.707 Hz. Justify correlation from actual common inputs and recorded conditions.

For example, the same proportional timebase error ε gives approximate readings fA(1+ε) and fB(1+ε). In the difference the common factor multiplies only the small true frequency difference. A 5 × 10⁻⁸ factor contributes 0.5 Hz at 10 MHz but around 3.5 × 10⁻⁸ Hz on a 0.7 Hz difference. This assumes the factor is the same for both. Trigger noise, cabling, temperature, and specimen changes still need allowance. Independent references with comparable calibrations need not have the same instantaneous error.

Keep paired observations

If the setup measures both channels at once from one reference, verify the channels with a common input, align gates, and save gate start and stop. After settling, record A, B, and their difference for every pair. Analyze the paired differences and keep raw readings. A shared upward trend can also be real temperature response in both specimens. Correlation in the traces alone does not establish correlated measurement error.

For a one-input instrument, alternate A, B, A and interpolate A to B's timestamp when changes are sufficiently linear. If the outer A readings are +1.0 and +1.4 Hz from nominal and the middle B is +1.9 Hz, the estimated difference is +0.7 Hz. Save timestamps and supply conditions. Steps, remaining warm-up, and changes in trigger settings can invalidate interpolation. Shorten the cycle, improve stabilization, or use simultaneous measurement where needed.

Keep absolute and paired judgments separate

Suppose the pair limit is 1.0 Hz and the observed difference is +0.7 Hz with standard uncertainty 0.316 Hz. With an agreed expansion factor k = 2, the interval is about +0.07 to +1.33 Hz and crosses the upper limit. Apply the established decision rule. If the individual limit is ±1.5 Hz, B's +1.9 Hz must be evaluated with its absolute uncertainty. A pair result does not change the individual frequency.

Record reference identity and calibration, gate settings, timelines, measurands, and covariance basis. Identify common and channel-specific contributions. Repeating can improve random estimates, but a fixed channel bias does not average out. Swap channels to check that bias. If the difference changes materially, investigate the measurement path before assigning it to the oscillators.

Verify common-reference and channel effects separately. Feed one stable source through a suitable splitter to check relative channel bias. Then restore the two devices and verify comparable amplitudes, slew, and loading. The observed difference still includes real short-term device motion, which can be part of the measurand if the true difference at a specified time is needed.

When changing benches, compare the same devices over several paired windows. A bench with one internal reference and one with separate references do not necessarily share the old correlation estimate.