A timing module uses a PCF2129 to log equipment events. On the production line a counter at CLKOUT reads near 32,768 Hz, and someone converts the reading into expected daily drift. Repeating the check on the same module at the 1 Hz output with an independent reference gives a different result. The RTC integrates a 32.768 kHz crystal and a temperature-compensated oscillator. Agree with the supplier on the output setting and test method before accepting a timekeeping specification.
Read the conditions below the default setting
NXP explains that the PCF2129 modulates its crystal load between two capacitance values using a programmable duty cycle. The device is factory calibrated and measures temperature at a selected interval. The manual recommends 16.384 kHz or lower at CLKOUT for an accurate frequency measurement. Its frequency characteristics also make specified RTC accuracy conditional on CLKOUT being set to a frequency other than 32.768 kHz or disabled. The default is useful as a quick indication of clock output activity. An accuracy measurement in that mode uses conditions different from the manufacturer's specified accuracy test.
Read the clock output control register at address 0Fh. The COF field sets 32.768 kHz at binary 000, 16.384 kHz at 001, 1 Hz at 110 and high impedance at 111. Preserve and record the temperature conversion field in the same register when changing the output. NXP lists a four-minute default interval and offers two minutes, one minute, or 30 seconds as alternatives. Allow the fixture and device to settle at each temperature, and capture a window appropriate to the temperature conversions. Checking a few cycles just after changing a setting produces results that are difficult to compare.
If a downstream circuit depends on CLKOUT, changing its setting changes that circuit's operating clock. Use a separate fixture or a mode where the downstream function can pause. If the deployed product needs continuous 32.768 kHz, independently qualify its timekeeping under that condition. The integrated 32.768 kHz TCXO is a different device and test object from a two-terminal passive crystal. A separate active oscillator also has its own test conditions.
Convert ppm into an observable quantity
At 16,384 Hz, a 2 ppm change is about 0.032768 Hz. At a nominal 1 Hz, it corresponds to about 2 µs per cycle or 0.1728 seconds per day. These are scaling examples. The product acceptance limit depends on the actual order code, temperature range and calibration conditions. The data sheet treats initial accuracy, temperature, voltage and aging as separate conditions.
With direct gated counting of a 16.384 kHz output, one count in a one-second gate represents approximately 61 ppm. One count in a 100-second gate represents about 0.61 ppm. Those numbers describe resolution only. Include reference and trigger uncertainty. A 1 Hz measurement can use a reciprocal counter or compare accumulated phase over many periods. Use an independent and calibrated timebase. Fixed propagation delay cancels in a difference measurement, but changing load and trigger thresholds still matter.
As another example, a 2.4 ms change in edge separation over twenty minutes against an independent seconds reference averages about 2 ppm. Record whether voltage and CLKOUT mode stayed constant. During a temperature ramp, the average describes that observation window. Accumulated edge phase is more informative at the ppm scale than a display rounded to whole seconds.
CLKOUT is open drain, so an external pull-up determines the rising waveform. Check that the actual waveform reaches the instrument threshold cleanly and has no extra triggers. With a hypothetical 10 kΩ pull-up and 100 pF total capacitance, the RC time constant is 1 µs and the 30% to 70% transition is about 0.85 µs. That number describes the test path, not the oscillator. A smaller pull-up increases sink current when low. Check the output limits and current budget, particularly under battery backup.
Distinguish inaccurate time from stopped time
Compare the RTC seconds register against an independent reference over a suitable interval. For whole-second jumps or interrupted timekeeping, record the stop flag, STOP setting and supply transitions before adjusting calibration. Some higher CLKOUT frequencies remain available with STOP asserted, so an output waveform does not prove that the calendar is advancing. Capture the main and backup supplies during switch-over when appropriate. Consider the documented aging correction only after confirming continuous timekeeping and a repeatable small error under the right test conditions.
Specify the full RTC order code, CLKOUT mode, voltage, temperature and reference on the procurement acceptance sheet. Preserve lot traceability and relevant reflow or storage conditions. Engineering should retain the fixture pull-up voltage, resistance and trigger settings. With that information, the team can distinguish a quick output check from a demonstrated timekeeping accuracy result.