Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

Why a low-noise oscillator trace can develop a deep notch as cross-correlation increases

Thousands of cross-correlations smooth a two-channel phase-noise trace, yet part of the curve dips. For an exceptionally clean source, investigate the measurement path. Cross spectra suppress uncorrelated receiver noise while preserving common contributions. If a contribution enters with opposite signs, it can cancel part of the DUT's positively correlated noise. NIST has documented this effect for thermal noise from power splitters and other measurement contributions.

Averaging can make the error repeatable

Use linear units. Suppose the desired real cross-spectrum component is +1.0 and an anticorrelated component is −0.8. The mean is +0.2, about 10 log₁₀(0.2) = −7.0 dB relative to the desired result. If the anticorrelated component is −1.2, the real part becomes −0.2. A plot based on cross-spectrum magnitude can show a similarly deep dip for either sign. Check whether the instrument exposes complex cross spectra, phase, or signed real values. The example assumes both contributions lie along the real axis and does not provide a correction for a real instrument.

Imagine one channel containing common signal s and disturbance n, and the other containing s and inverted n, with the two random processes independent. Cross-spectral averaging retains the spectrum of s minus that of n. More averages narrow the spread around that difference. A repeatable trace therefore can still contain systematic anticorrelation. Positive correlated bias may also raise a trace.

Change the measurement path

Record the original wiring and receiver input powers, splitter, buffers, isolators, filters, source load, and instrument options. Export single-channel noise data and the cross spectrum where available. Run a verified check source with the same settings. Correct receiver compression or insufficient input level before comparison. Confirm offset frequency, integration range, and whether the detector is sensitive to AM, PM, or both. AM leakage of opposite sign in phase detector channels can also cause cancellation.

Repeat using a characterized splitter or an independent method supported by the equipment. Hold input power, terminations, and reference plane comparable. A different splitter may also change actual oscillator load, so record it as a variable. High isolation alone does not establish negligible anticorrelated thermal noise. NIST reported that splitter type and added isolation or amplification could affect results near the thermal floor. If another adequate setup is unavailable, report the affected offset region and distinguish independently reproduced results.

Decide whether the notch supports acceptance

Suppose the limit is −160 dBc/Hz at 100 kHz offset and the first trace reads −165 dBc/Hz in the suspect dip. That number alone does not establish conformance. A verified alternate setup reading −158 dBc/Hz under matched conditions should be evaluated with its uncertainty. If both paths leave the dip unresolved, retain frequency ranges and original traces rather than drawing through the affected region. Specify which bands support integrated results and which need further evidence.

If several units show similar dips, investigate the shared configuration. See how position and depth change with the splitter arrangement. Reducing the correlation count can reveal statistical scatter, but that scatter does not recover the missing quantity. Fix the qualified setup and check it before testing another lot. Save phase information where the analyzer reports negative real values.

Preserve cross-spectrum phase where possible. Where a nearly pure anticorrelated term dominates, the signed real component can become negative and phase can approach 180°. A magnitude image shows only a dip. Channel group delay and instrument reference also affect phase, so check sign with known common modulation before interpreting it. Noisy single-channel traces still include receiver contributions.

If the tool saves only plots, log offsets on either side of the dip, readings in each setup, correlation time, and environment. Independently check the measurement structure before ranking oscillators. A requirement at one affected offset cannot be answered by substituting neighboring valid points.

References

WeChat

MERLIN · TIMING COMPONENTS

Scan with WeChat to contact Sales

Add our WeCom contacts from your usual WeChat app.

On desktop, scan with your phone. On mobile, save the QR image and select it from your album in WeChat Scan.