A time interval counter shows a nearly constant 2 ns difference between two timing modules. Reconnecting a termination changes that difference to 1 ns. Save the voltage waveforms and determine the voltage at which the counter recognizes each edge. Changes in amplitude and slew move the time of a fixed-threshold crossing. That movement can matter to the actual receiver. When comparing the timing sources internally, establish which parts of the observed shift come from the crossing definition, cables, and inputs.
How two edges starting together can appear apart
Consider linear rising edges. Channel A goes from 0 to 2.0 V in 8 ns, so it rises at 0.25 V/ns. Channel B goes from 0 to 1.0 V in 8 ns, or 0.125 V/ns. The edges start at the same instant. With both thresholds at 0.5 V, A crosses in 2 ns and B in 4 ns, producing a 2 ns apparent delay. With individual thresholds at their 50% levels, 1.0 V and 0.5 V, both cross in 4 ns. If both absolute thresholds are 0.75 V, the apparent difference becomes 3 ns. The direction of change in a threshold sweep is diagnostic.
Real 1 PPS signals may have offset, overshoot, reflections, or compression. Calculate percentage thresholds from the observed low and settled high levels. Changes in termination can change waveform shape, too. A percentage comparison aligns the selected crossing point. Retain the fixed receiver threshold where the interface requires one.
Feed both channels from one source, then restore the real inputs
Use a common pulse source within the load and voltage limits of the setup. Match the cables and terminations used in the actual measurement. Set both inputs to the same trigger level and measure the residual skew. The splitter and cables can also contribute. Reversing inputs while tracking the original sign helps distinguish channel and path effects. Keep splitter loading consistent. For a high-impedance CMOS output, use the proper termination or a suitable buffer. Changing input impedance can change voltage and reflections as well as the indicated crossing time.
Reconnect the actual pulses and capture both edges. Record slew near each crossing and sweep through several valid trigger thresholds. Take multiple pulses at each setting and record input impedance, polarity, and coupling. If the measured difference follows the direction predicted from local slopes, the thresholds and waveforms contribute. If it remains substantially constant, investigate channel delay, cable propagation, and real source offset. When exchanging inputs, watch for counters that always measure from a start to the next stop. If the edges reverse order, the reading can gain a full second. Use signed two-channel records or a known, correctable offset to preserve edge order.
Define acceptance at the signal source and receiver
Assume an example tolerance of ±1 ns for B relative to A. A fixed 0.5 V crossing at the actual receiver yields a 2 ns delay in the example, beyond that limit. A separate comparison at each source's own specified half-amplitude crossing finds alignment under that definition. Save both measurement planes and crossing rules, along with how the levels were determined, how thresholds were checked, and the observed variation. Repeated agreement establishes repeatability, which can coexist with a constant bias.
Check the instrument's threshold resolution and accuracy near a decision limit. At 0.25 V/ns, 20 mV corresponds to 80 ps for A. At 0.125 V/ns, it corresponds to 160 ps for B. Allow for these contributions in the measurement uncertainty before judging a marginal result.
Define the statistic as well. The mean interval estimates relative bias, while individual excursions can reveal intermittent slow edges, interference, or multiple counts. A consistent 2 ns and 4 ns time to threshold can produce a very narrow distribution and still leave a 2 ns bias. Link waveform captures and interval values by pulse epoch. If reflections create two crossings, changing counter hysteresis is useful only when the waveform also establishes which crossing is counted. Define how abnormal crossings are handled before screening.
Before relying on automatic triggering, identify the threshold it selects at each amplitude. An algorithm based on observed peak-to-peak voltage can respond to noise, pulse duty cycle, and time window. For cross-bench comparisons, save each channel’s numerical threshold. Identical percentage labels can still correspond to different voltage settings.