During a continuous phase comparison on an OCXO, the device cable passes through a temperature chamber while the reference cable stays outside. Halfway through the heating ramp, the phase plot starts to slope. Stop the ramp and the slope fades. Check the oscillator’s temperature sensitivity and the cable’s propagation delay. A fixed length mainly changes arrival time by a constant amount. Delay changing from second to second adds a phase slope. Converting that slope to frequency places a cable effect in the device reading. NIST distinguishes fixed path delay from changes in delay in its material on frequency transfer.
Define the phase sign and units
Use a positive time difference for arrival later than the reference. If propagation delay increases, each marker arrives slightly later relative to the one before it. The period inferred from arrival times increases and frequency falls. Divide the change in delay by elapsed seconds and reverse the sign to estimate fractional frequency error from this effect. Match the convention to the instrument’s channel order. Some report device minus reference and others use reference minus device. Copying another laboratory’s sign can double the temperature correction. Try a noticeably longer cable as a fixed-length control to confirm the sign in the log. After it stabilizes, the longer cable mainly adds a constant phase offset.
Worked example with assumed values. A stable 10 MHz source travels through a cable whose delay increases from 50 ns to 51 ns between samples 100 s apart. The delay rate is 1 ns / 100 s = 10⁻¹⁰. The apparent fractional frequency error from the cable is approximately −0.1 ppb, or about −0.001 Hz at 10 MHz. Identical changes in both cables can largely cancel in a differential comparison. Different positions inside the chamber, bend radii or materials can leave a residual. Determine the real delay change by comparing the paths. Also check how the phase algorithm handles wrapping across a complete cycle after the assembly is connected.
Use a single source as a control for both channels
Before testing, split one stable signal through the full device path and the full reference path. Keep the planned splitter, connectors, feedthrough and cable placement. Terminate both as required and record the splitter’s conditions at the measurement frequency. The input frequencies are then identical, so differences characterize the paths and instrument residual. Take a baseline with the chamber idle, then run the same temperature profile. Put ambient temperature in the chamber, temperature near the cable and the measured time difference on one time axis. The chamber setpoint describes the program. The cable itself usually changes temperature later.
If possible, swap the two cables and check whether the temperature-dependent phase trace follows one path. A swap reverses differential comparison order, so label it using the established sign. A loose connector or moved cable can leave a sudden step. Log door openings, fixture adjustments and reconnections. Keeping the path fixed helps distinguish a reconnection offset from a continuous slope during temperature change. To evaluate a more stable cable, compare it in place. Vendor temperature characteristics need the actual cable construction and temperature span.
Save the raw time series with the chamber plot
Use the control to determine whether cable drift can affect interpretation of the limit. If it is far below the required resolution, retain the control with the method record. If it approaches the apparent device change, consider routing both cables through similar temperature zones, shortening the exposed length or moving the room-temperature split closer to the fixture. Reducing the error is easier to qualify than subtracting an old trace. If correction is needed, document repeatability of temperature, cable routing, termination and instrument state, and report the residual. Save raw measurements and the control used for correction.
Chamber fans can vibrate both oscillator and cable. A control can mount the cable in the same way while feeding it from a stable source outside. If changing only the path moves the feature in the trace, examine cable and fixture. If the common-source measurement is quiet and the sample follows airflow, distinguish airflow on the package, mechanical vibration at its support and supply changes. Use consistent input bandwidth and stable input power on both channels. Cable loss may vary with temperature and turn a level change into timing error in the detector.
If the sample cannot be removed, reserve a spare channel with comparable cable as an environmental witness. Disclose the positional difference and avoid treating it as an exact correction. When splitting one source, check that reflections at one port do not feed the other arm.
At chamber plateaus, check that both paths settled and compare the time needed for the device and cables to reach thermal equilibrium. A supplier’s hours-long plot can be checked by asking whether a common-source control covered the full temperature range and whether the time-to-frequency conversion used a defined channel sign. Fixed cable delay and how quickly that delay changes during the test are separate quantities.