Extra lines can appear when a scope is switched to its highest real-time rate with only one input enabled. Some scopes reach that rate by alternating ADCs. Small lane differences in offset, gain, or timing create spectral components tied to the sampling cadence. Change both sample rate and input frequency to see whether the suspect line follows that cadence before declaring oscillator spurs.
Where the extra components land
In a two-way interleaver, one ADC takes samples 0, 2, and 4 and the other takes 1, 3, and 5. At 200 MS/s the odd-even pattern repeats every 10 ns. An offset mismatch creates a component near 100 MHz. Gain mismatch alternately multiplies samples by slightly different factors. For a 25 MHz input, the simplified two-way model places an image at 100−25 = 75 MHz, with the other sideband folding in the sampled spectrum. Alternating time skew can produce an image at the same frequency. Here it coincides with the oscillator's third harmonic.
If lane gains differ from the mean by +0.005 and −0.005, ignoring other mismatch, the simple alternating model gives an image amplitude ratio of about 0.005, or 20 log₁₀(0.005) ≈ −46.0 dBc. That estimates detectability. Actual instruments may interleave more converters, correct digitally, or have other timing and frequency-response effects that prevent identifying a cause from one line alone.
Separate the predicted image from a harmonic
Substitute a low-distortion sine source with the same sample rate, vertical range, termination, and bandwidth. Changing the source from 25 to 24 MHz shifts the predicted image from 75 to 76 MHz while moving the third harmonic down from 75 to 72 MHz. This can separate components that previously coincided. Verify the replacement source's own spurs. Record shared-reference lock and acquisition settings if used.
Then keep the oscillator frequency fixed and select a different documented sampling or channel configuration. At 180 MS/s in the same two-way topology, a 25 MHz signal yields a predicted image at 65 MHz. Read the actual rate. Changes in sweep or record length may affect only stored data while the ADC continues using its original sampling architecture. Document the hardware mode separately from the spacing in the exported record.
Verify with another path
Cross-check 72, 75, and 76 MHz on an independent spectrum analyzer if available. Keep the load seen by the oscillator consistent when switching or splitting paths. Check input power and attainable floor using a clean reference. Keep the scope's FFT window, record length, and amplitude normalization consistent. A longer record or a window with lower sidelobes helps separate closely spaced components and carrier leakage. Resolve coincident lines before assigning the marker reading.
If raw records expose the lane order, compare even and odd means or fit each subsequence to a known sine model. Ensure adequate input filtering and actual raw samples. Exported resampled or interpolated points need not map to physical lanes. Repeatable frequency shifts, mode changes, and independent measurements are stronger evidence.
Record the acceptance conditions
Suppose the third harmonic limit is −50 dBc and the 75 MHz line on the fastest scope setting reads −46 dBc. Investigate the overlap before disposition. If another acquisition moves the suspect line and an independent analyzer measures the 75 MHz third harmonic at −58 dBc, evaluate −58 dBc with its uncertainty. Keep carrier frequency, actual sample rate, channel topology, and independent trace together. Save measurement bandwidth and detectability limits as well.
A second check uses two suitable input amplitudes at the same frequency. A gain-mismatch image follows the carrier in the linear range, while an offset mismatch component can appear with the input terminated and no signal applied. Keep the front-end range and coupling comparable and avoid inadvertently switching attenuation or preamplification. This amplitude check helps separate a fixed offset pattern from an input-dependent image.