Some boards occasionally miss an edge at the first power-up and recover after reset. Acquisition averaging smooths a repeated waveform by averaging different acquisitions at corresponding time points. A fault in a single record is attenuated. If the start times vary between acquisitions, averaging can even produce a slow transition that never occurred in any individual run.
Capture the actual event first
Trigger on the supply or a dedicated enable signal. Use single acquisition with enough pretrigger history to capture a true cold start. To estimate how often the fault occurs, save each raw record and count failures before using averaging to look for regular interference. If only one of sixteen acquisitions is missing a high pulse and the other fifteen are normal, the average remains close to the normal high level.
Some scopes have a high-resolution mode that combines adjacent samples within one acquisition. It can be used on a single event, unlike averaging across triggers, but it can smooth nearby details in time and change effective bandwidth. To verify a very narrow gap, save the standard acquisition first and confirm sample interval and instrument bandwidth before interpreting a high-resolution trace.
Include the failing trace and a normal trace under the same conditions. Record the power-off duration before each cold start. If an average is also included, state the number of acquisitions, trigger alignment, and purpose. Where the software can replay individual acquisitions, use the power-event number to find the failing record and measure its amplitude or duration directly.Tektronix oscilloscope acquisition modes