Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

What limits a measured 3.5 ns edge on a CMOS oscillator output?

A CMOS oscillator has a steady frequency reading, but the reported rise time is longer than expected. Identify the probe, input channel and any bandwidth limit used. A setup that separates successive cycles clearly can still be too slow for the shortest edge within a cycle.

Start with the fastest edge the oscilloscope can display

For an approximately Gaussian measurement response, estimate instrument rise time as 0.35 divided by bandwidth. At an effective bandwidth of 100 MHz, that is about 3.5 ns. Other response shapes can use different coefficients, so check the specific instrument documentation. Even a much faster signal edge appears longer in this setup. Directly subtracting the displayed time from the datasheet maximum does not recover the device's own rise time.

A device's datasheet may measure from 20% to 80% of its swing while the instrument defaults to 10% to 90%. The latter spans more of the transition, so even sufficient bandwidth does not make the readings directly comparable. Apply the datasheet thresholds under the same load. If overshoot is significant, also check how the instrument establishes the high and low plateaus and selects crossings.

Change one condition at a time when repeating the measurement

Save the raw waveform and bandwidth settings. First change channel bandwidth using the same probe, measurement point and thresholds. If the measured rise time shortens markedly with greater bandwidth, the original setup contributed to the measurement. Repeat with a probe of known bandwidth and input capacitance, and record how it changes loading. Photograph any changed ground accessory and connection point. If the edge is still too slow with the faster probe, engineering can investigate fitted loads and routing and identify the waveform suitable for the supplier.Tektronix guidance on oscilloscope bandwidth, rise time and sample rate