The first Si570 reaches the target frequency, so its register values are baked into firmware. The next prototype then shows a consistent offset. The manufacturer's FAQ explains that each internal crystal has its own actual frequency. A value derived from the first part does not apply to later parts. Calling the lot defective can reject usable parts while leaving the faulty assumption in production firmware.
First establish which unit supplied the tuning constants
Follow the datasheet to calculate the internal crystal frequency from each unit's own startup configuration and known factory output frequency. Use that result to calculate the target registers. Save the raw register inputs and note whether they were read before or after the first software frequency change. Registers left over from the preceding run can give a wrong starting point. Record device position and traceability identifier to avoid combining inputs from different boards.
Use a cross-check before rejecting the lot
Measure both boards using settings calculated from their own parameters, then reproduce the old fixed-constant method under controlled conditions. Check whether the error follows the calculation. Use the same traceable reference and record warm-up, output load and measurement resolution. If a unit still fails after per-device calculation, ask about its accuracy option, supply and actual test conditions, and retain the raw records.
Production test software should read and calculate per unit as part of the normal process. A debug file that happened to pass is unsuitable as a common factory preset. In the customer report, distinguish the part conditions, firmware version and passing board identifiers so the acceptance test remains repeatable after replacing the same model. Si570 and Si571 manufacturer guidance on per-device calibration