A lab uses two sine-wave sources at the same frequency to simulate reference clocks for a low-input-power board test. After aligning their phase, the operator reduces power on one output. The displayed phase suddenly jumps. The board may be sensitive to amplitude, or the source may switch its internal path when power is adjusted. Remove the board from the signal path first. Keep the generators, coax and matched terminations, then measure phase at the board input plane at every planned power setting. If the jump also occurs with the calibration load in place, that measurement provides a baseline for the test.
A power adjustment can change the components in the signal path
Some RF generators use an electronic step attenuator to cover a broad power range, with a variable attenuator for smaller changes within each coarse range. The step attenuator contains switches and multiple attenuation sections. Switching ranges can change the combination of switches and components the signal traverses, changing phase and propagation time. Rohde & Schwarz application note 1GP108 explains this in the RF level section. Switching points can depend on the calibration of the individual instrument. Even two generators of the same model, set to the same indicated output power, should have their actual switching points checked separately. A common 10 MHz reference can tie the source frequencies together. Measure offsets from attenuator path changes separately.
Keep nominal frequency and the commanded phase fixed. Let both paths warm up, then leave one source unchanged as a reference. Change only the RF output level of the other. Route both through the cables, adapters and terminations intended for the test into an appropriate two-channel instrument. Keep channel range, coupling and input impedance consistent, and verify the lowest level can still be resolved. A small signal can move the trigger crossing as its voltage changes. With a scope that has sufficient amplitude resolution and avoids clipping, fit the two sine-wave phases or use zero crossings with the same waveform center and direction. Separate changes in level-dependent crossing time from changes in actual carrier phase.
Three readings show how to apply the baseline
Worked example with assumed values. Set both sine waves to 100 MHz, giving a 10 ns period. Define the result as the lag of signal 2 relative to signal 1. In a through fixture, assume the lag changes from 6° to 18° when output goes from −2 dBm to −8 dBm. The extra 12° in the same direction represents 10 ns × 12/360, or about 333 ps at that frequency. Suppose the board measurements are 9° and 21° at the two power settings. Subtract each setting’s through baseline at the same measurement plane. Both cases give 3°, or about 83 ps. These numbers illustrate the calculation. Check that the baseline applies if there are board reflections, overload or substantial errors across ranges. A single carrier phase also leaves an ambiguity of whole periods. Use the time reference required by the project for broadband pulse arrival measurements.
Run the current power setting, one lower setting, and the original setting again. Repeat in the upward direction. If the phase step repeats at the same power and the phase recovers on returning to the original setting, keep a separate baseline for each relevant range. If the switching point moves or the restored setting has drifted, check the reference, warm-up, connectors and automatic input ranging. If there is a third monitoring channel, log actual source power to verify comparable conditions. Change only source attenuation during each comparison. Keep the receiving instrument on its original range. If changing that range is unavoidable, use a stable signal to check whether the new range introduces a phase offset. Also record the wait from a power adjustment to data capture, to separate the switching transient from steady-state behavior.
To reuse a baseline across shifts, save the generator serial number, external reference connection, carrier frequency and measured power with the instrument setup. After replacing or calibrating equipment, take a through measurement at each end of the operating range. Even if the instrument change has no measurable effect, this confirms that the phase reference plane still agrees with the report.
Choose the adjustment method before sweeping power
Some instruments can hold the step attenuator fixed and use the remaining adjustment range to change output power. Some support digital attenuation with an applicable baseband configuration. A fixed range has limits. Digital attenuation changes signal-to-noise margin, so check whether phase is still measurable at the lowest power. An external attenuator brings its own phase, connectors and impedance. Measure it at the same reference plane after inserting it. The manufacturer separates the fixed-range option from recalibration of phase at different levels. For a sweep of just a few decibels, first verify that one fixed range covers it. If the test must switch ranges, retain a baseline for each formal test setting and verify repeatability in both sweep directions.
A power change can also expose an actual delay change on the board. For example, a fixed comparator threshold away from the sine-wave center produces a different crossing time as amplitude changes. Keep the sine-wave phase measured at the board input and compare it with the output edge. If the input phase jump repeats with generator settings, use the verified range-specific baseline. If input phase stays stable and only board output moves, investigate threshold, bias and input loading. Give receiving inspection the generator ID, power, cables, baseline and device output as separate fields. Purchasing can request both the input and output traces for a borderline low-power result and find the affected stage before deciding whether to replace the oscillator or receiving board.