A record marking VDD as 2 V can look like an undervoltage test of the SiT9121. A specific SiT6097EB LVPECL setup shifts the device’s supply reference so the instrument ground provides the desired termination reference. Figure 4 shows a 3.3 V setup with VDD at +2 V and device VSS at −1.3 V relative to instrument ground. The voltage across the device remains 3.3 V.
Use this only with the manufacturer’s specified split-ground configuration and compatible parts. A board that ties DUT ground to the SMA shell cannot simply reuse the same supply settings. With power off, verify the configuration resistors, ground nets, and both supply connections. Check the 50 Ω instrument input and DC coupling against the drawing. Keep instrument protective earth connected.
On power-up, measure the device’s positive rail relative to its own ground, then measure device ground relative to instrument ground. Record the reference used for each single-ended output voltage. A second lab using conventional biased termination may observe different absolute levels. Matching frequency does not establish the same DC load.
For acceptance at the customer’s receiver, perform final validation with the customer board’s bias and input topology. The split-ground setup reproduces the manufacturer’s DC-coupled instrument condition. Include the full wiring and measured voltage across the device with quotations or return records. Distinguish instrument ground from DUT ground directly on the diagram to prevent a misleading interpretation of “2 V.”SiTime SiT6097EB manual, Section 3.1.4 and Figure 4