After a precision meter is connected to J4, the 32.768 kHz clock continues but the current reading is unexpectedly low. Trace R3 in the evaluation board schematic. J4 is intended for device current measurement. If the default zero-ohm link is still fitted, current can bypass the meter, leaving very little in the parallel path. Determine whether the device supply current actually passes through the meter shunt.
With power off, configure the measurement path according to the manual and check the meter range, fuse, and lead sockets. Once the meter bridges the open link position, measure the actual voltage at the device supply pin and then observe the clock. Burden voltage on a sensitive current range can reduce the device voltage, particularly during autoranging or startup. Loss of output under those conditions alone does not establish a defective oscillator.
The manual illustrates a reading under one load. For comparison with a datasheet, state the exact part, output voltage option, load, and buffer path. Exclude the board amplifier’s supply current from the oscillator’s microamp reading. Use the same range to compare light load and the intended receiver. If the reading is unusually low, first check that the bypass is open.
When handing the board to another person, record whether R3 has been reinstalled. If it oscillates with the series meter in place and stops when the meter is removed, the supply path may simply be open. Keep both assembly states on record. If readings jump when the range changes, repeat on a suitable fixed range and confirm that the actual supply stays within device limits.SiTime SiT6098EBB manual, Sections 3.4 and 4.2.5