Comparing short-term stability of two TCXOs: why specify averaging time in the RFQ?

Purchasing receives two very small short-term stability values, but the designer wants to know how much the local clock varies when the equipment corrects frequency every 10 s. Find the averaging time in the specification row, often denoted by τ. Allan deviation (ADEV) describes how the average frequency changes between adjacent intervals. The same device can give different values at different averaging times. Keeping only the smallest value can misrepresent an optimum observation interval as a specification for all intervals.

Calculate one point from a frequency record

For the ordinary nonoverlapping calculation, split a continuous record into equal 10 s intervals. Average each interval and express its frequency relative to nominal. Find the difference of each adjacent pair, square the differences and average them. Divide by two and take the square root. If three 10 s averages are +2, +4 and +2 ppb, the two differences are +2 and −2 ppb, and ADEV is about 1.41 ppb. This short sequence illustrates the calculation. An evaluation requires enough intervals to characterize the result.

Adding the same fixed offset to all three readings leaves their differences unchanged. A part that runs consistently high can still have good short-term stability. If the equipment also needs calibration, specify accuracy relative to nominal separately. Mark interruptions or missing measurements in the raw record. Joining samples across a gap as if they were adjacent changes the meaning of the calculation.

Keep the column headings when reading the datasheet

The SiTime SiT5543 datasheet, for instance, lists the ADEV value with a 10 s averaging time in the typical column. Ask suppliers for the value at the same averaging time, whether it is typical or guaranteed, the measurement temperature and whether drift was removed. If equipment also corrects once per second, request the 1 s value or the full curve. Each mode can then be evaluated without guessing a curve from one 10 s point.Allan deviation row in the SiTime SiT5543 datasheet