A time-interval instrument must measure both narrow pulses and long delays. Changing the display units is insufficient when using a TDC7200. The longer-interval mode combines a count of full external reference periods with fine measurements at each end. The other mode addresses shorter intervals. Choose from the expected minimum and maximum interval and the actual reference period.
A 16 MHz reference has a 62.5 ns period. Intervals of tens of microseconds span many full reference periods. If software omits the clock count and converts only the fine timing, the result can be far too short. For very brief inputs, forcing the long-interval mode can also produce an unusable result. Check the mode-specific limits, reference frequency, and minimum number of reference periods in the manual.
Use a pulse source with known intervals to test one short and one long case. Check the register reads and calibration conversion in each mode. Capture ENABLE and CLOCK together so you can see whether the external reference is stable and remains present while enabled. A stale reference period can make results from an intermittent clock look deceptively precise.
Attach mode, raw counts, calibration status, and reference period to each result. Define whether the instrument reports out-of-range inputs as invalid or changes range. One valid calibration point does not establish the full measurement range. If both the clock and mode are changing, test the new algorithm with the old reference before switching clocks.TI TDC7200 measurement modes, counters, and CLOCK requirements