In pulse ranging or industrial flow measurements, crosstalk after transmission can reach STOP before the real echo. In the applicable clock-counting mode, the TDC7200 can ignore early STOP events through an initial mask. This window is defined in reference-clock counts. Recalculate it before carrying over a register value to another reference frequency.
Suppose the report specifies a blanking interval of 80 reference cycles. At 8 MHz, each cycle is 125 ns and the resulting interval is about 10 µs. Raising the reference to 16 MHz with the same count shortens it to about 5 µs. Verify the exact boundaries against the IC’s comparison and qualifying-edge definitions. Extending blanking can suppress early crosstalk, but it can also discard valid echoes at the shortest range.
Use a two-pulse source to generate a fixed early interference pulse and an adjustable valid STOP. Observe the physical input and the reported stop time. Move the valid pulse across the mask boundary and record the first position where it is accepted consistently. Repeat without the interference pulse to distinguish mask behavior from comparator saturation or pulse amplitude differences.
If a measurement requires several valid STOP events, verify the expected count, overflow flags, and incomplete status. This avoids reading stale data from the prior measurement as the first echo. Record the reference frequency, mask value, and measured interval so minimum range can be rechecked after a crystal or firmware change. Near the mask boundary, move echo arrival time while keeping amplitude fixed to avoid confusing threshold variation with mask-count error.TI TDC7200 CLOCK Counter STOP Mask register