A programmed oscillator reads correctly on a counter, but TimeFlash 2 reports a large deviation. The manufacturer says the measurement panel computes ppm relative to the frequency shown in the programming panel and applies the selected threshold. A leftover target from a previous batch can produce a precise-looking failure against the wrong frequency.
If the intended target is 25.000 MHz but the interface still shows 25.001 MHz, an actual reading of 25.000 MHz will be displayed as approximately −40 ppm. Recalculate the error from the raw reading, the target, and the formula before testing the sample again. Correct the comparison reference before touching the device configuration. Some supported devices use one-time programming, making reprogramming an inappropriate diagnostic step.
After checking the displayed target, save the threshold and its basis. A wider debug limit does not change the rated stability. Incoming room-temperature limits also do not replace temperature and aging specifications. For a borderline part, retest with a traceable reference at the specified supply and load and keep both readings.
Save the measurement log before exiting, together with the programming-panel screenshot and actual part number. If changing the displayed target clears the failure, record a configuration discrepancy. If the raw frequency still differs from the correct target, investigate socket contact, instrument timebase, and the sample. This sequence preserves the sample for later analysis.Microchip TimeFlash 2 manual, Section 3.4 and Figure 3-4