A service technician measures 6.78 MHz on a TRF7970A board and requests a crystal at that frequency. Start by identifying the test point. On pin 27, SYS_CLK is a buffered output for other circuitry. A 13.56 MHz crystal divided by two and a 27.12 MHz crystal divided by four both produce 6.78 MHz there. The two hardware options require different crystal orders even though that counter reading is identical.
Identify the net and read the active clock settings
TI's TRF7970A Rev. M data sheet assigns SYS_CLK to pin 27. The adjacent DATA_CLK pin 26 is an interface data-clock input. Trace the pad and confirm which clock the host MCU actually uses. A similar label or a one-pin probe placement error changes what the measurement means. Keep the test-point identity with the PCB revision and any clock source selected inside the MCU.
While EN is high, read the Modulator and SYS_CLK Control register at 0x09. B7 enables the 27.12 MHz crystal option, and B5:B4 are Clo1:Clo0. Table 6-36 gives the following mapping. The bits describe the setting, so establish which source is fitted through the part marking, approval drawing or an appropriate independent measurement.
| B5:B4 | Divider | 13.56 MHz reference | 27.12 MHz reference |
|---|---|---|---|
| 00 | Disabled | No SYS_CLK | No SYS_CLK |
| 01 | Divide by four | 3.39 MHz | 6.78 MHz |
| 10 | Divide by two | 6.78 MHz | 13.56 MHz |
| 11 | Divide by one | 13.56 MHz | 27.12 MHz |
If you have verified a stable 6.78 MHz on pin 27, a divider setting of 10 implies a 13.56 MHz reference and 01 implies 27.12 MHz. Check that B7 agrees with the inferred configuration. If the readout and the measured frequency disagree, check for a downstream divider, a moving firmware state or a mismatch in the test-point identification. It helps to associate a register capture and output waveform with the same operating stage. TI TRF7970A Rev. M, register 0x09 and the SYS_CLK selection table.
Distinguish startup and sleep readings from the normal output
With EN low and EN2 high in Power Down Mode 2, the chip switches SYS_CLK to a typical 60 kHz auxiliary clock. The period is about 16.7 µs. At 6.78 MHz it is about 147.5 ns. Label the active and auxiliary portions on a startup capture so a slow reading during wakeup is interpreted in the right state. The data sheet gives a typical startup time on the order of 3.5 ms from full power-down. Use the product's timing requirement for acceptance.
The register table describes reset behavior after power-on reset, EN low and writes to ISO Control, with a stated exception for the divider bits on ISO Control writes. Record readbacks after initialization, after a protocol change and after resume. Register 0x09 also controls modulation. An arbitrary whole-byte write to test a clock frequency could change RF operation. If the MCU itself depends on SYS_CLK, first establish an appropriate supported test setup before changing the divider. TI TRF7970A Rev. M, oscillator section and auxiliary clock.
Order against the actual reference circuit
The passive crystal goes between OSC_OUT and OSC_IN, pins 30 and 31. Table 6-9 specifies the recommended 13.56 or 27.12 MHz fundamental-mode, parallel-resonant crystal characteristics. It lists ±20 ppm frequency tolerance, aging less than 5 ppm per year and a −40 to +85 °C operating range. If the original board uses an active clock source, the data sheet shows an external input on pin 31 with pin 30 left open. Follow the supply, ground and output connections to identify the circuit rather than identifying the part by its package alone. A crystal order should include its exact frequency, load, ESR limit and pad drawing. A clock oscillator order should also verify drive levels, supply and enable behavior.
When both load capacitors are equal, TI's first estimate is CL ≈ C1 ÷ 2 + CS, where CS accounts for the chip and PCB stray capacitance. Its 18 pF and 4.5 pF example gives two 27 pF capacitors. A hypothetical substitute requiring 12 pF would start with two 15 pF capacitors if the 4.5 pF estimate still applies. Leaving in the two 27 pF values would still estimate 18 pF, 6 pF above the hypothetical candidate's requirement. Verify loading, startup and frequency on the actual assembly. TI TRF7970A Rev. M, crystal connections, load example and recommendations.
Frequency tolerance also cannot distinguish the two paths at one output. In an ideal divider example, ±20 ppm gives ±271.2 Hz on 13.56 MHz and ±542.4 Hz on 27.12 MHz. Divide those errors by two or four respectively and both contribute ±135.6 Hz at 6.78 MHz. This calculation isolates the initial input tolerance. It leaves the crystal’s temperature, loading and aging effects to their own budget. For an order approval, specify the reference tolerance under the relevant fixture conditions and log the SYS_CLK observation separately.
For acceptance, confirm the reference source and clock configuration independently, then test the host. Prefer a known buffered measurement point because a probe on a crystal pin can change its load. As a timing check, suppose a factory diagnostic waits 6,780 MCU cycles and uses 6.78 MHz directly. The wait is 1.00 ms. If the host mistakenly receives 13.56 MHz with the same counter settings, the wait becomes 0.50 ms. Trace the real MCU clock tree before blaming the crystal or retuning a delay. Test the product's startup, wake and normal communication modes using the approved test tag. Put 6.78 MHz in the measured-output field of the service record, and use the verified reference frequency and full part number to request a replacement.