A board runs clocks at 100 and 101 MHz. The analyzer sees 99 MHz only when both clocks are enabled. This fits third-order intermodulation. The product could arise in a shared board amplifier, combiner, or connector, or after the analyzer's input. Keep the board operating state fixed while varying the analyzer mixer level. Track both carriers and suspect lines.
Predict both nearby products
The near-carrier third-order products are 2 × 100−101 = 99 MHz and 2 × 101−100 = 102 MHz. Each is 1 MHz away from the nearer carrier. Check both sides of the full trace for distinguishable components. Filtering, path gain, and floor can make the two levels different.
In the weakly nonlinear third-order model, reducing both inputs to a mixer by 10 dB lowers the internally generated product by about 30 dB. Relative to either carrier, that is about 20 dB. An analyzer typically refers its display to the input port after adjusting internal attenuation. Its carrier readings should stay similar while internal intermodulation changes. Suppose both carriers read −15 dBm and 99 MHz reads −65 dBm, or −50 dBc. With 10 dB added input attenuation, an internally generated product alone would read about −85 dBm, or −70 dBc. A product already present at the input should stay close to its original corrected input level if the new setting leaves matching and compression essentially unchanged.
Change only attenuation after the board
Record 99, 100, 101, and 102 MHz with the board mode, bandwidth, preamp, and other carriers fixed. Change internal attenuation manually by several steps and verify consistent corrected main tones. Address any compression or overload indications. Some instruments change paths with reference level or attenuation. Log path settings and change one variable at a time. Applying a pad externally requires a separate loss correction.
Repeat with a second low-distortion path. Where sources feed an external combiner, check source isolation and any hidden nonlinear device such as a limiter or active probe. Verify each source into a suitable matched load before combining. If an unexpected line appears only when combining, substitute a characterized combiner or isolate the sources appropriately. To investigate intermodulation in a shared board amplifier, vary levels upstream of it in a separate experiment with a stable measurement chain.
Report detectability when the line falls into the floor
If the floor at the second setting is −82 dBm, a predicted −85 dBm product may no longer be distinguishable. Record it as unresolved under those conditions. A −65 dBc limit for a −15 dBm carrier requires assessing power below −80 dBm. A −82 dBm floor leaves little measurement margin. Improve sensitivity or use a verified lower-distortion instrument. A narrower RBW can lower noise, provided the sweep settles and carrier skirts are understood.
The real and internally generated products may add at the same frequency. Increasing attenuation can produce nonmonotonic changes. Avoid subtracting levels from two captures as if the simple sum were known. Seek a range of attenuation settings with stable results within the usable dynamic range, then corroborate with another path. Preserve both source settings, spectra, matching information, and verified detection bounds for any unresolved result.
Frequency provides a second experimental variable. Moving only the second clock from 101 to 101.1 MHz should move the ideal products to 98.9 and 102.2 MHz. If 99 MHz stays fixed, look for another fixed source or a reference-related spur. The movement checks the combination but does not locate its origin. Use it alongside the attenuation test. For fixed-frequency oscillators, a substitute two-tone source at similar powers and frequencies can check the analyzer. Label substitute-source and device results separately.