A control accelerometer sits at a specified point on a shaker or fixture. The oscillator may sit at the edge of a flexible PCB supported by standoffs. At some frequencies a 1 g table input can produce 2.4 g near the oscillator. Normalizing clock response only to the table then incorporates board and fixture amplification. For system acceptance, the installed assembly's response remains relevant because that is what the receiver experiences.
Measure input and local response
In a sinusoidal sweep example, the table indicates 1.00 g peak at 300 Hz while the response point reads 2.40 g peak. The amplitude ratio is 2.4 or about +7.60 dB. At 450 Hz the local response might be 0.70 g for the same input, or −3.10 dB. Use consistent peak, RMS, or peak-to-peak values. For one sinusoid, 2.40 g peak is about 1.70 g RMS. Capture phase as well. A substantial phase lag can indicate a mechanical mode and invalidates pointwise multiplication of the table trace by 2.4.
Suppose the clock shows ±1.2 ppb modulation at 300 Hz. Dividing by the table reading gives 1.2 ppb/g. Dividing by 2.40 g measured nearby gives 0.50 ppb/g for the particular installation, axis, and frequency. Board flex, cross-axis motion, and cable microphonics may also contribute. Document mechanical boundaries and frequency before interpreting the result as device sensitivity.
Account for the mass of the response sensor
Pick a location near the part that avoids its pads and routing. Check accelerometer mass, mounting, and usable response. If an unmodified assembly has a feature near 300 Hz and adding a heavy accelerometer or magnetic base moves it to 270 Hz, the instrumented response may differ from the original condition. Use a suitable low-mass sensor and verified mounting, or an appropriate noncontact method. Soft mounts can attenuate high frequencies, and heavy additions can shift resonances.
Acquire control and response channels simultaneously and record axes and mounting photos. Verify antialias filtering and amplitude and phase response near the top of the band. Dress cables so they do not brace or pull on the PCB. Even with single-axis input, local transverse motion can be significant. Label the result accordingly.
Separate fixture modes from installed assembly modes
Try a permitted low-level survey of the empty fixture, then repeat with the board fitted. A peak already present in the empty fixture warrants checking the fixture and control location. A peak that appears only after installation warrants checking support, enclosure, and connectors. Record original conditions before changing standoffs or torque to locate the effect. For acceptance under real mounting, return to the required installation. Document sweep rate and dwell near narrow peaks because the response may not reach steady state during a fast pass.
If the requirement is clock continuity under 1 g at the table, control that input and check edge anomalies, frequency excursions, and unlocks with the required assembly. If the task is device response estimation, document local acceleration, board flex, and reference conditions separately. Preserve both acceleration traces, mounting photos, timing records, and fixture changes so a later assembly can be compared.
Use the same interface, fasteners, and baseplate when comparing lots. Keep empty and loaded fixture sweeps. A different socket or a stiff temporary coax can change how force reaches the oscillator, so verify the local response remains comparable. If the corner is too small for a sensor, map the gradient at nearby points on representative boards and report the distance between response point and device.
Align acceleration and timing acquisitions on a common timeline. With separate instruments, use a common trigger and verify latency stability. Compare the 300 Hz acceleration and phase components before interpreting transfer. If changing cable restraint changes the phase curve, investigate measurement leads before drawing conclusions about lot consistency.