Consider a sample report that includes a stability plot for a 10 MHz oven-controlled crystal oscillator (OCXO) labelled “one-second frequency samples.” The counter actually measures for one second, then spends another second retrieving, saving and retriggering. Setting the Allan deviation software to a one-second interval would treat measurements with start times two seconds apart as adjacent one-second windows. Even a smooth result cannot directly verify a drawing limit on σy(1 s). Inspect the acquisition timeline before judging the smallest value on the plot.
Separate the counter’s two time intervals
One interval is the measurement window τ0. The other, T, is the spacing between successive window starts. A conventional contiguous series of rectangular frequency averages requires both to be equal. If the first average covers 0–1 s and the second covers 1–2 s, the windows meet. If the first covers 0–1 s and the second covers 2–3 s, any frequency variation from 1 to 2 s is absent from the data. That uncovered interval is dead time. When the host computer writes one row every two seconds, determine whether the instrument actually measures at two-second intervals or acquires continuously and transfers batches from a buffer. The file write time alone cannot establish whether there is dead time.
Relabelling the horizontal axis as two seconds does not repair the example. Each value still averages only one second of the signal, and the missing second is still unknown. Filling the gap with the previous value artificially suppresses some differences. Linear interpolation introduces correlations. The dead-time section of NIST SP 1065 explains that corrections depend on the noise type and sampling pattern. If the purchase specification has no agreed correction model, the most direct retest is to acquire a contiguous series and deliver it with the original timestamps.
Calculate four readings by hand to see what is being compared
Take an illustrative series of four contiguous, dead-time-free, one-second rectangular averages. Relative to nominal 10 MHz, the measured offsets are +0.007, +0.009, +0.008 and +0.012 Hz. Divide by nominal frequency to obtain fractional frequency y of 7, 9, 8 and 12 in units of 10−10. The adjacent differences are 2, −1 and 4, giving squared differences of 4, 1 and 16. The two-sample Allan variance divides their sum by twice the number of terms. Thus σy(1 s) = √[(4+1+16)/6] × 10−10 ≈ 1.87 × 10−10. Adding a fixed frequency offset to all readings leaves those differences unchanged. Even excellent stability therefore calls for a separate check of the mean frequency against calibration or incoming inspection limits.
At two seconds, the four points give only one complete pair of adjacent two-second averages, 8 and 10. The calculation yields about 1.41 × 10−10. One pair demonstrates the arithmetic and cannot establish convincing acceptance confidence. An actual report should include the raw record length, the number of usable difference terms at each averaging time and, where possible, the equivalent degrees of freedom or confidence intervals. Overlapping estimators produce more difference terms, but those terms share data and cannot be counted as independent samples. Avoid ranking suppliers by a single low point at the far right of the plot.
Bench retest sequence
- Specify the measurand.Confirm whether the required statistic is ordinary Allan deviation (ADEV), its overlapping estimator (OADEV) or modified Allan deviation (MDEV), which adds another averaging operation. Record the specified averaging time and measurement weighting. If computing externally from counter frequency readings, check the weighting of each reading. Equally spaced phase or time-error samples can be used with the NIST second-difference formula for OADEV. Record the instrument mode and software version for either method.
- Measure the baseline before connecting the sample.Use a signal that meets input and loading requirements. Splitting a single source into two channels to measure relative phase mainly characterises the differential measurement chain. Some common reference noise cancels, so that test alone cannot establish that the reference is more stable than the oscillator under test. Verify the reference separately at the target averaging times. Record cables, termination, trigger settings and environment to help trace differences when samples are swapped.
- Acquire continuously inside the instrument.The Keysight 53230A application note, for example, demonstrates gap-free measurements to investigate close-in stability. Verify the exact mode, timestamps and transfer procedure on the actual setup. A measurement study by Dunker and coauthors reported temporal correlations in the 53230A’s continuous, high-resolution output. Gap-free acquisition and measurement weighting are separate checks. The mode name alone cannot establish that the output is equivalent to independent rectangular averages.
- Reproduce the operating conditions.Describe the warm-up before recording, target temperature and actual fluctuations, supply and load, external reference and treatment of drift or outliers. To qualify a device, keep the original curve and include a separate detrended curve if the method was agreed in advance. If the equipment has to tolerate drift in service, include the raw behaviour in the judgement so processing does not hide changes that matter in the field.
Request two sample qualification files
The first is the raw phase or frequency series with timestamps, window length, start spacing, missing-sample flags and continuous acquisition mode. The second is a report listing each target τ, the stability estimate, confidence information, the reference and the measurement floor. When there are occasional missing points, split the data into contiguous segments first and exclude differences spanning the gaps. If producing a combined estimate, document the statistical method and applicable conditions for each segment. When the target τ exceeds what the contiguous record can reliably support, a retest is more productive than stretching the curve.
If the report retains only a frequency column, request the trigger settings and the reason for each wait in the acquisition script, then cross-check a short record. For instance, the first ten acquisitions may cover every interval while a range change drops one second. Instrument timestamps or status markers can identify that interruption. Where possible, repeat the check at the beginning and end to detect buffer overflow or pauses that develop only during long acquisitions.
If two suppliers provided only PDFs labelled “one-second averaging” and “one-second sampling,” ask them to report the same estimator at the same averaging times with verifiably contiguous data. The technical agreement can specify separate requirements for stability at the chosen τ values, mean frequency offset and the measurement floor. Each incoming report value then maps to a defined requirement, and retesting can distinguish the device, reference and acquisition process.