Two delivery reports for quartz oscillators say the seal passed. One provides a helium mass spectrometer reading and fine-leak test conditions. The other supplies a single value labelled equivalent air leak rate. Supplier A has the smaller number. Does that establish a tighter package? First identify the seal that was tested, the range of leaks the method covers and the measurand each number describes. A numerical ranking can be wrong even when the units look identical.
Mark the package boundary on the report
A passive quartz resonator cavity and the external housing of an active oscillator are different physical boundaries. An oscillator may have a sealed resonator and separately packaged oscillator circuitry. The report may cover the crystal, the complete oscillator or a process witness with a similar seam. Ask the supplier to mark the actual boundary on the assembly drawing, identify the joining method and state whether the cavity was backfilled or evacuated. NDK’s precautions for crystal units explain that vacuum or inert gas packaging helps maintain crystal performance and that corrosion can impair sealing. The manufacturer’s construction identifies which interface matters. A metal lid is a clue to look up the construction drawing.
Some test settings depend on free internal gas volume. This is the space gas can occupy in the sealed cavity. External length multiplied by width and height is a different quantity. For example, a 7 × 5 × 2 mm package has an outer bounding volume of 70 mm³. If its free volume is 8 mm³, that is 0.008 cm³, nearly nine times smaller than the 0.070 cm³ bounding volume. Ask the manufacturer for the volume or volume range. Recheck the applicable method and limits when the package shrinks, the construction changes or more solid material is added inside. An old test limit may no longer apply under the same conditions.
Identify the measurand before converting units
MIL-STD-883-1 Method 1014.19 distinguishes the tracer-gas leak rate indicated by a detector under specified test conditions from an equivalent standard air leak rate stated for reference conditions. If both suppliers use the same method and measurand but express the result in different units, convert units directly. Since 1 atm = 101,325 Pa and 1 cm³ = 10−6 m³, 5 × 10−8 atm·cm³/s = 5.07 × 10−9 Pa·m³/s ≈ 5.07 × 10−8 mbar·L/s. The gas and meaning of the measurement carry through. Only the unit changes.
If one value is the instrument’s indicated helium throughput and the other an equivalent air leak rate, the arithmetic above is insufficient. In addition to gas properties, conversion depends on the conditions defined by the method. When the calculation needs the helium exposure pressure and duration, the dwell between removal from the pressure chamber and measurement, and free volume, losing one field can make the result irreproducible. Check whether pressure is absolute or gauge. Some gas factors in the method apply only under specific conditions, so a fixed multiplier cannot convert arbitrary tests. Ask the manufacturer to retain the detector reading, reported measurand, units, conversion method and the conditions that justify it.
Why a clean fine-leak trace needs gross-leak coverage
For a fine-leak test using helium pressurisation, tracer gas enters through the leak and escapes during detection. A package with a large leak can lose much of its tracer during transfer or waiting, which can give a low fine-leak reading. Small volumes and different dwell times alter the detectable range. An experimental study of miniature packages tested the same sample at different dwell times and showed why a single apparent reading in a small cavity can misrepresent leak severity. The study used MEMS packages and illustrates the importance of the measurement window. Its dimensions and results are separate from the oscillator specification.
A report may list fine and gross leak separately or cover both with a combined method. Check coverage against the revision, conditions and flow specified in the purchase order. As one manufacturer example, the Q-Tech B Plus screening specification lists fine and gross leak in separate rows. A note gives the corresponding test conditions and a helium fine-leak value of 5 × 10−8 atm·cm³/s. It also states a condition for a particular small-volume case. The number illustrates how to read a real document. For a different series or purchase using another revision, consult the applicable document and test report before carrying an older condition identifier into a new inspection specification.
Review the incoming report against the supporting evidence
- Match the complete order code, revision, cavity, screening option and measured lot. A qualification report and a lot test report support different conclusions. The latter should indicate full inspection or sampling and the scope of the sample.
- Check fine- and gross-leak conditions, detection limits, and instrument or calibrated leak verification. Treat a below-detection result as an upper bound. If the detection limit exceeds the purchase requirement, request a result capable of resolving that limit.
- For a borderline result or missing gross-leak coverage, retain the supplier report, affected units and same-lot controls. Let the responsible lab review under the agreed procedure. The purchase document and test method determine whether retesting can be used for acceptance.
When changing suppliers and comparing different benches, consider a two-site comparison with representative check units. Use the same identifiers on units and test records. Where the method permits, compare calibrated leak certificates, empty chamber background and the detectable range under common conditions. Check restrictions on repeat runs, transport and specimen condition. Use separate check units of the same construction when necessary. Because marginal results depend on the test window, recording sequence and environment helps trace differences to cavity assumptions, test settings or bench background.
Record seal screening and electrical performance separately. A device can meet incoming frequency limits while its seal is in doubt. A device with a passing seal record can still miss long-term frequency stability. Preserve the tested boundary, method revision, decision measurand and detection range when comparing suppliers. This identifies the evidence supporting release and the points to revisit on a future package change.