Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Passive Crystals and Oscillator Circuits

Estimating crystal motional capacitance from the spacing between an impedance notch and peak

A lab sends a sweep of a 10 MHz crystal with a dip on the left and a peak on the right. Two people select markers from different traces, convert the spacing into motional capacitance and obtain different answers. Start with the measured quantity, the calibration plane and the origin of each feature. The spacing can provide a useful initial model when those details are known. Check that model against another measurement at a known load.

Identify series, antiresonance and load resonance separately

In a low-loss model of one mode, the reactances of the motional inductance L1 and motional capacitance C1 cancel at the motional series resonance fs. At a slightly higher frequency, the branch is inductive. Its susceptance can cancel that of the crystal's shunt capacitance C0, producing high impedance at antiresonance fa. Applying an external parallel capacitance CL produces a separate load resonance fL. The ordered frequency may be specified at such a load. Keep the three frequencies separate in the test report and say how each marker was found.

A calibrated impedance-magnitude trace ordinarily shows the main mode as a low-impedance dip followed by a high-impedance peak. An admittance or transmission trace has a different response and needs its own interpretation. With finite loss, an impedance minimum, a zero-phase crossing and the motional series frequency can differ slightly. Use peak and dip locations as estimates for a clean, low-loss mode. For a model that other engineers will use, fit the complex data, state the instrument convention and check the residual. Keysight's equivalent analysis reports motional series resonance separately from its resonance and antiresonance markers. NDK equivalent circuit and electrical characteristics · Keysight frequency definitions in four-element equivalent analysis

Extract the motional capacitance from a 20 kHz separation

The lossless one-mode model gives fa = fs√(1 + C1/C0). Rearranging gives C1 = C0[(fa/fs)² − 1]. For a hypothetical crystal, use a series frequency of 10,000,000 Hz, an antiresonance of 10,020,000 Hz and independently established shunt capacitance of 3.00 pF. The result is 0.012012 pF, or 12.012 fF. In the same model, L1 = 1/[(2πfs)²C1], which gives approximately 21.09 mH. These figures illustrate the calculation. The precision of an actual result depends on the setup. Analog Devices discussion of crystal impedance and antiresonance

Input change in the hypothetical modelResultCheck before using it
Two markers and the device's 3.00 pF12.012 fF motional capacitanceUse matched marker definitions and shunt-capacitance data
The fixture adds 0.25 pF, but the calculation still uses 3.00 pFApparent antiresonance near 10,018,463 Hz and apparent motional capacitance near 11.09 fFRemove the fixture contribution
Substitute the 12 pF load frequency for series resonanceLoad frequency near 10,004,003 Hz produces a spurious estimate near 9.60 fFUse the series frequency for extraction and the load frequency as a check

The added fixture capacitance raises the total shunt capacitance to 3.25 pF and draws the two characteristic frequencies closer together. Omitting the correction makes the estimate about 7.69% low in this example. The external 12 pF case instead uses the load resonance with the known external capacitance added to the device capacitance. This is a useful check on the extracted model. Apply a known load to a real sample and compare its observed frequency with the prediction. If they disagree, check drive, load definition and the assumed single mode. If an adjacent spurious response affects the fit, evaluate that branch separately or ask the manufacturer for an extraction method. Epson equivalent circuit and load-frequency relationship

A close pair of markers makes uncertainty easier to overlook

Try a second hypothetical example. A series frequency of 10,000,000 Hz, an antiresonance of 10,000,600 Hz and a shunt capacitance of 3.00 pF give about 0.360 fF. Suppose the two marker frequencies each have a ±20 Hz error bound. At opposite extremes the separation can be 560 to 640 Hz. Allow the shunt capacitance to vary between 2.90 and 3.10 pF and the corresponding input extremes give about 0.325 to 0.397 fF. Reporting extra decimal places hides this spread. With actual data, account for correlations such as a common frequency reference. The example treats the marker error bounds as able to move in opposite directions.

Start with a sweep wide enough to see the neighboring responses, then narrow the span. Preserve the complex impedance or data from which it can be reconstructed. Record calibration plane, fixture correction, crystal drive, frequency step, temperature and repeat mounting. A source setting in dBm does not determine the power dissipated by the crystal. Use the fixture method and resonant impedance to establish the drive seen by the device. Keysight's series-through example uses open, short and load impedance calibration at the fixture. If the antiresonant peak is close to the instrument's high-impedance limit, check measurement range and resolution before deciding whether to use independent shunt-capacitance data or a complex model fit. Keysight crystal series-through measurement and calibration

Pass on the trace and its purpose with the extracted value

Ask the supplier for the full code, vibration mode, method and conditions, and explain whether the result is for simulation, qualification or incoming inspection. A fitted sample value can inform a simulation if its source and residual are documented. Retain the raw trace and the result at a known load with it. If a purchasing requirement calls for a guaranteed limit across lots, agree on that separately. For a related discussion, see the MERLIN guide to motional quality factor and fixture-loaded bandwidth.

WeChat

MERLIN · TIMING COMPONENTS

Scan with WeChat to contact Sales

Add our WeCom contacts from your usual WeChat app.

On desktop, scan with your phone. On mobile, save the QR image and select it from your album in WeChat Scan.