Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

Does a deep notch in delay-line phase-noise testing mean the oscillator is especially quiet?

Some phase-noise tests split one source. One branch goes straight to a comparator, while the other takes longer to arrive. The comparator sees present phase minus earlier phase. If a periodic phase disturbance makes exactly one full turn during the delay, it has the same value in both paths and the difference is small. The measurement chain loses sensitivity near such an offset and exhibits a response null. Check delay and correction before interpreting a notch as part of the oscillator spectrum.

Both comparator inputs come from one source

An analog delay line and detector, or an analyzer using different digital delays, compares phases separated by a known time. For slow variations the two observations tend to be close. At higher disturbance frequencies they can differ more. When the disturbance period equals the delay, the two observations line up again. Subtracting a delayed phase waveform gives a magnitude response of 2 × |sin(π × offset frequency × delay)|. Phase power-spectrum correction uses the square of the amplitude factor and the detector or demodulator calibration.

Dividing measured power by a tiny factor close to a null magnifies floor, measurement error and resolution limitations. A correction cannot reconstruct phase variation that did not register. Analog arrangements also need a proper carrier operating point in the detector. A longer delay can increase loss, so verify both input levels. If the instrument stitches bands or selects delays automatically, check how it marks valid offsets and whether exported traces are already corrected. Keysight diagrams the analog delay-line and digital discriminator techniques separately.

Worked example with assumed values. A demonstration chain has an effective delay of 1 µs. For a 10 MHz carrier, that is ten carrier cycles, but the null depends on the phase fluctuation offset. At 1 MHz offset, the disturbance makes one cycle in 1 µs, giving zero ideal difference. At 500 kHz offset, it advances half a cycle, giving an amplitude factor of 2. Reducing the delay to 0.8 µs moves the first nonzero-frequency null to 1.25 MHz. The amplitude factor is then about 1.18 at 1 MHz, so the ideal chain responds there. These values illustrate the calculation and do not specify any instrument. Sweep offset with a known phase-modulated signal within the allowed modulation depth, using two delays and checking whether the notch moves as predicted. Check carrier and modulation limits before relying on the automated report.

Determine whether the feature moves with delay

Save the original trace, instrument delay and applied power near a notch. Reacquire with only delay changed, keeping load, supply and reference conditions the same. If the notch moves inversely with delay, first treat it as a measurement-chain feature. Check critical offsets with another phase comparator or instrument whose principle is understood. A dip or protrusion at the same offset then warrants checking the oscillator, regulator and nearby circuitry. Temperature and carrier drift can move an instrument’s operating point, so save state and settling time for each run.

Document which offset bands can be compared. Some settings have poor close-in sensitivity. At larger offsets there can be multiple nulls or front-end bandwidth limits. Compare different settings in overlapping bands with adequate margin to validate correction and units. Without a second measurement covering a notch, state that sensitivity is inadequate there and identify the measured bands. Do not fill the gap with low values for jitter integration. When combining two traces, mark crossover points and their respective configurations.

Amplitude variation can also create false phase if an analog comparator is off its operating point. Check a known PM stimulus at several offsets before testing the device. Record both response away from nulls and actual level. If cable-length changes shift power, first set each arrangement to its specified input conditions and then compare. A response null moves to the location predicted by delay. A change that follows only power calls for checking gain, floor and operating point. In the close-in band, look for slow warm-up drift. Stabilize the source first and keep the approximate null positions from bench calibration.

When comparing low-noise devices, purchasing can request raw results at critical offsets with the measurement method. Retest units near a limit with another valid configuration if the limit falls near a null. Discrete lines, spectral density and integrated jitter are related. Trace each reported result to a band with adequate sensitivity before using it for a purchase or replacement decision.

References

WeChat

MERLIN · TIMING COMPONENTS

Scan with WeChat to contact Sales

Add our WeCom contacts from your usual WeChat app.

On desktop, scan with your phone. On mobile, save the QR image and select it from your album in WeChat Scan.