When measuring a differential oscillator output near the receiver, a small pulse sometimes appears in step with a power switch. Both board signals may be disturbed together. To check how much common-mode signal the instrument puts into the differential result, connect both probe tips to one low-voltage test signal within the input ratings. At the probe inputs, the ideal difference is zero. If a residual appears, record its size and shape and check the probe’s common-mode rejection at the relevant frequency. Return to the normal probe points for the actual board measurement. Do not short independent outputs together.
The differential front end has two slightly different paths
Differential voltage is the subtraction of two node voltages. Equal simultaneous disturbances should cancel. Input capacitance, connection length, tip accessories and front-end response can differ slightly between the two probe sides, changing the amplitude and timing after the signal travels each path. Tektronix defines common-mode rejection ratio, normally reported in decibels, and shows how it varies with frequency. Strong rejection at low frequency does not predict fast-edge performance. Many digital waveforms repeat at tens of megahertz while their edges contain much higher frequencies. For a spike at the onset of an edge, consult the specified rejection across the relevant band.
Another option is subtracting two single-ended channels. Each has its own gain, delay and quantization error. A small differential voltage on top of a large common-mode signal may require a wide front-end range. Subsequent subtraction cannot recover the small detail lost before digitization. Moving the math trace to the center of the screen will not restore it either. A true differential front end with sufficient common-mode range can amplify after subtraction. Still check the differential range, common-mode range, maximum input rating and tip accessories.
Worked example with assumed values. Apply the same 0.8 V peak-to-peak sine signal to both inputs, keeping DC bias and instantaneous values within limits. If the residual at that frequency is 8 mV peak-to-peak, the ratio is 0.8 V / 0.008 V = 100. The equivalent rejection is 20 times the base-10 logarithm of that voltage ratio, or 40 dB. At a lower frequency, a 0.8 mV residual would give about 60 dB. These numbers illustrate calculation. Compare the actual signal level, frequency and floor. A residual at the noise floor supports a lower bound under those conditions. Trace thickness alone cannot establish an exact rejection value.
Set up a zero-difference check
Use a properly terminated line to drive a low-impedance node on a small test fixture. Connect both tips to that point using the manufacturer’s accessories. Keep the probe attachments symmetric and use a ground and supply reference representative of the board. Pick an allowable amplitude and a sine frequency representative of the interference. Then use a suitable transient signal to observe fast edges. Long leads can change common-mode behavior. Keep the fixture short and verify the local signal amplitude. Turn off unrelated measurement functions and save attenuation, offset, bandwidth, coupling and source settings. The result characterizes the complete probing path. Check the manufacturer’s load, accessories and definitions when comparing with its specification.
If the real differential sources have unequal impedance, rejection may degrade on the board even when the probe does well on a low-impedance fixture. Reproduce representative impedances or capacitances in a low-risk substitute fixture and repeat the common-mode test. If the original tip leads differed in length, compare again with matched accessories. Reversing the probe polarity should invert the genuine differential waveform. Whether the residual holds its amplitude or changes sign is an additional clue to interpret along with the same-node test.
Another check is to run the board under the affected conditions and move both tips to the same nearby node. If it is not low impedance, connecting both tips adds load. Start with the correctly loaded fixture, then use a low-load probe on the real circuit. If interference reaches the two lines at different times, a true board-level conversion from common mode to differential mode can appear. The same-node control cannot cancel that real voltage difference. Save both single-ended traces relative to local ground, deskew the channels and compare edge order. Use a dedicated front end if dynamic range is limited. Record whether the result comes from a fixture residual or a true board asymmetry. Capture a no-signal baseline on the same range before the check.
Compare the millivolt-scale board anomaly with the residual from the control. If they are comparable, treat it as a measurement capability limit and try a better differential front end or shorter connections. If the control residual is small and a clear pulse remains under load, check the two series elements, differential termination and receiver supply. For incoming oscillator verification, provide the estimated common-mode error and the same-node trace with the measured anomaly.