A DMTD measurement reads 30 microseconds. What is the 10 MHz RF time offset?

Two 10 MHz oven-controlled crystal oscillators go into a time interval counter for direct RF edge comparison. The readings fluctuate. After switching to dual-mixer time difference, the two zero crossings are tens of microseconds apart. Before using the record to qualify a sample for a frequency standard, determine whether those microseconds occurred at RF or were magnified in the mixer beats. Treating beat time directly as RF time can misstate the result by six orders of magnitude.

Distinguish two 10 MHz inputs from the common local oscillator

Dual-mixer time difference (DMTD) puts device A and reference B into separate mixers, both driven by the same slightly offset transfer oscillator. Filter out the high-frequency sum terms and retain two low-frequency beats that are close in frequency. For example, A and B are nominally 10,000,000 Hz and their common oscillator is 9,999,990 Hz. Both beats are near 10 Hz. The input phase difference is present in the beats, but at the lower frequency one radian spans much more time than it does at RF.

The NIST dual-mixer time difference diagram and equations show how the beat zero crossings drive a time interval counter. This article uses a sign convention suitable for acceptance. With the transfer oscillator below the RF signals, the reading is positive if the device beat’s positive-going zero crossing is later than the reference crossing. Ignoring channel offsets, the device RF arrival delay x is approximately the beat time difference Δt multiplied by fb/f0. A high-side oscillator, opposite crossing polarity or a different start-stop connection can change the sign. Verify it with a known delay. If the two beat frequencies differ slightly, use the measured reference beat frequency and make sure that it is much greater than the frequency difference between device and reference.

Calculate an example before choosing the instrument

For a 10 Hz beat on a 10 MHz carrier, the time magnification f0/fb is 1,000,000. A +30 µs beat reading corresponds to +30 ps late at RF after baseline correction. A second reading +50 µs at a point 0.20 s later corresponds to +50 ps. With late time positive, average fractional frequency relative to the reference is approximately −(50 − 30) ps/0.20 s = −1 × 10−10, or −0.001 Hz at 10 MHz. A slower device accumulates lag, consistent with the negative frequency sign. Absolute arrival difference requires correction for fixed measurement-path offset. A constant path offset cancels in the time slope used for frequency difference.

A 1 ns time interval counter resolution scales to 1 fs in the ideal ratio. That describes only the quantisation step. Usable DMTD capability also depends on mixing, filtering, zero-crossing circuits and transfer oscillator routing. A nearly sinusoidal 10 Hz beat with 1 V peak-to-peak has a zero-crossing slew of approximately 2π × 10 × 0.5 = 31.4 V/s. With 0.1 mV of threshold offset near zero, the crossing can move by about 3.18 µs, equivalent to about 3.18 ps at RF. The calculation shows why the circuits ahead of the counter may limit actual measurements. Verify the system floor and offset experimentally.

Three checks for scale, offset and cycle slips

  • Apply a known phase change.Split one stable source into two paths and add a known small delay to one. Verify the sign and scale. Nominally, a 100 ps RF delay changes the 10 Hz beat time by approximately 100 µs. If the bench can repeat at a 20 Hz beat, the unscaled change should roughly halve and the RF-equivalent result should agree. Include the uncertainty of the calibration delay or the method used to establish it.
  • Run a common-source baseline.Split one signal into two measurement branches and record beat amplitude, threshold, temperature and relative time difference. Check different relative phases. A single test with nearly coincident zero crossings covers a special case. Common transfer-oscillator noise can cancel in ideal balanced paths. Mismatched filters or sampling at different times can leave residual contributions. The NIST frequency measurement chapter describes DMTD, its low-pass filters and shared oscillator. Reconnect an independent stable reference and establish that its fluctuations do not dominate the sample result.
  • Record beat periods and acquisition status.If the counter reports a time difference within one beat period, wrapping near zero can create a jump of almost 0.1 s. At this magnification that is about 100 ns at RF, one 10 MHz cycle. If more than one cycle can elapse between valid samples, use additional evidence to resolve how many were lost. Software can unwrap phase only under stated continuity and maximum-rate conditions. Filtering every large jump can hide both missing counter measurements and actual device loss of lock.

Choose the beat frequency according to how fast a change must be seen. One sample per zero crossing of a 10 Hz beat gives about ten points per second. Reducing the beat to 1 Hz increases time magnification tenfold but extends the sampling interval to about one second. To capture a change lasting tens of milliseconds, use acquisition with the required time resolution or a higher beat frequency and verify the mixer and filter path at that setting. Report both the observed beat frequency and the final output sample rate. Gating, averaging or missed triggers can make them different.

For absolute delay, establish channel phase offsets at a defined reference plane. At 10 MHz, a constant 1° phase difference corresponds to 100 ns/360, or about 278 ps, potentially larger than the acceptance limit. Measure the offset with a common-source baseline, connect the sample without changing cabling or filters, and verify that the baseline remains stable. If reconnecting changes the result by tens of picoseconds, include connection repeatability rather than applying the first correction to every configuration.

A qualification report should name both sources, the common oscillator frequency, measured beat frequency, sign convention, sampling period, floor and baseline treatment. If two setups produce different curves, compare them with the same source and a known delay before attaching the samples. That avoids judging devices while changing the measurement chain at the same time. For a stability limit at a specified averaging time, apply the same estimator to the time records and retain the number of valid samples and measurement floor. DMTD makes a sensitive relative phase measurement. Absolute device accuracy still depends on the reference and calibration chain.