The same oscillator measures differently after reinsertion. Is it the fixture or the sample?

Two shifts suddenly report different yields even though the daily standard-source check passes. Select stable samples that can be repeatedly mounted and have both shifts test the same units. A standard source connected directly checks the counter and part of the cabling. It usually bypasses production sockets, contacts, mounting and device power. A crossed test with actual samples measures variation at one station and differences after changing the operator or fixture. It can start with retained units.

Define repeatability and reproducibility in terms of the actual procedure

Repeatability describes the variation when measuring the same unit again under the same conditions. Reproducibility addresses changes across operators, instruments or time. The NIST gauge study guidance covers design and analysis, including samples, operators, repetition and order. Fixture, supply and counter all participate here. If routine inspection requires reinsertion, repeated trials should include remounting. Reading ten values without moving the unit mainly shows readout variation.

An oscillator may drift during its first seconds after startup or respond to hand heat and seating pressure. Standardize condition before insertion, power-on delay and output load. Use the same written procedure across operators. A common order works for a preliminary check. For a formal study, randomize sample order and blind previous results so warm-up and repeat order do not coincide. Have each operator measure all representative samples with at least two independent remounts. One operator can also run both fixtures to help separate procedure and equipment. Log concurrent contact repairs, compensation changes, probe changes and reference status.

Worked example with assumed values. Operator A reinserts a retained sample twice in a new fixture and reads +4 and +6 Hz, averaging +5 Hz with a 2 Hz difference. Operator B measures +11 and +13 Hz in an old fixture, averaging +12 Hz. The means are 7 Hz apart, greater than either within-group difference. Swap fixtures while keeping procedures and repeat both trials. If both operators read high with the old fixture, the result follows the fixture. If operator B reads high with both, investigate their procedure or station. One device and two readings do not establish formal repeatability and reproducibility. The example designs the next test and is not a disposition rule.

Look for contacts, waiting time and fixture loading

If the discrepancy appears only on new lots, cross-test retained units of the same nominal frequency and suspect units, and check the variation against the relevant specifications. Worn contacts may change supply and return. Measure voltage at the device. A passive crystal socket and probe loading can change actual load capacitance, so verify electrical conditions at the fixture. Where scripts convert several nominal frequencies to ppm, check from raw hertz. Keep unrounded readings. Separate reinsertion tests from uninterrupted repeat readings to distinguish contacts and readout scatter.

A study may show that one operator is more consistent for one package, for example because seating force differs with package height. A single large retained unit cannot qualify all smaller packages. Include the actual part families, frequency offsets and sockets. If the setup can read input and output, record local voltage or load with results. Keep order numbers when variation concentrates at first insertion. They may reveal contact, thermal settling or power-up sequence. After a fix repeat the same sample set and retain before-and-after traces.

For a blind rerun, renumber units from different lots and use the same insertion, seating and delay procedure. An operator who knows a unit was borderline might take additional readings before reporting it. Decide repeat count and treatment of first failure ahead of time. If a group cannot repeat under the original conditions, log the change separately from earlier repeatability. Preserve original order to see sustained drift. A shared witness unit at beginning and end adds another baseline check.

Report enough evidence to define the rework scope

Record same-condition variation and differences across operators and fixtures, then decide which results near a limit need retesting. Use an adequate sample matrix for formal gauge statistics and consider product tolerance and use. Calibration or a comparison may still be needed to detect a common bias. Purchasing should be able to trace each row by part number, sample, fixture revision and operator. Changes across all three indicate where to investigate. If the difference follows only the device, ask the supplier to repeat under the same test conditions.