Equally spaced peaks appear on both sides of the clock carrier, and the supply ripple has the same frequency. That coincidence suggests possible coupling, but the switching node, other periodic loads or the probe ground can all introduce the same tone. First record the ripple frequency, the spur spacing from the carrier and the spur amplitude separately.
Try a test that shifts the suspect frequency
Within the device's allowed operating modes, change only the suspect regulator's switching frequency. Keep the clock configuration, load, instrument reference, spectrum analyzer resolution bandwidth and attenuation unchanged. If the ripple frequency and spur spacing move together, they have a common frequency source. Further tests are needed to distinguish coupling through supply pins, radiated coupling and probe pickup.
Restore the original frequency, then change one load or compare with a suitable low-noise supply. Check probe grounding and the measurement points. Retain paired readings of frequency and amplitude before and after each change. If you propose a filter change, check regulator loop stability and supply droop at startup again before approving it.