A designer changes one nominally 50 Ω cable for another on a 100 MHz low-phase-noise OCXO, and the counter shows a different frequency. If the data sheet specifies frequency change under load variation, first establish the receiver's complex impedance as seen at the oscillator, then test whether reflections pull the oscillator. The 50 Ω markings identify the cables’ nominal characteristic impedance. When the receiver is mismatched, changing cable length rotates the reflection phase and can change the load seen by the oscillator.
Two data sheet entries serve different purposes
Page 2 of the KVG O-22-ELPN-100 MHz data sheet specifies a sine-wave output and a 50 Ω ±10% load. It separately lists frequency stability versus supply variation and versus a 50 Ω ±10% load variation. The load-related limit is ±10 ppb, equivalent to ±1 Hz at 100 MHz. Purchasing should distinguish the load the output expects from the permitted frequency change under the stated load variation. If applying this specification to a real complex impedance or a switched load, send the actual load trajectory to the manufacturer for confirmation.
Output loading sensitivity characterises the complete oscillator. It differs from the picofarad load capacitance in a passive crystal data sheet. Output reflections can couple through the buffer into the oscillator loop. Changes in output driver current can also couple through the supply. A different fixture can truly shift frequency, or it can make the instrument read differently because signal amplitude, noise or triggering changed. An isolated frequency difference supports a conclusion about the device only after the measurement chain has been checked.
Similar return loss can produce different frequency shifts
With a 50 Ω reference, the reflection coefficient is Γ = (ZL − 50)/(ZL + 50). Begin with purely resistive examples. A 45 Ω load gives Γ ≈ −0.0526 and return loss of about 25.6 dB. A 55 Ω load gives Γ ≈ +0.0476 and about 26.4 dB. The reflected magnitudes are similar and the phases are opposite. A 50 + j10 Ω load gives a magnitude of about 0.0995, return loss of about 20.0 dB and VSWR of about 1.22. Chapter 2 of the Keysight network analyser guide gives the relationships among reflection coefficient, return loss and VSWR. A report showing only return loss cannot reproduce the corresponding complex impedance.
When a low-loss cable connects a mismatched load, the reflected wave makes a round trip. The reflection phase at the oscillator equals the load reflection phase minus twice the one-way phase shift. At 100 MHz, a cable with a propagation velocity of 0.66 times the speed of light has a wavelength of about 1.98 m. Adding 0.25 m changes the round-trip shift by about 91°. If two lengths give different counter readings, change only that section and keep the connectors, supply and counter setup. With a truly matched receiver at 100 MHz, changing the length of an ideal cable alone does not create significant reflection.
Start with a precision load, then reproduce the customer’s load
- Define the reference plane.Calibrate the network analyser at the connector or board reference point the oscillator will see. Measure the downstream network, powering the receiver in its relevant state where necessary. Protect the instrument from voltage or RF power at an active port. This step measures receiver-network reflection in each state, along with adapters and actual cable lengths. Do not connect an oscillating output directly to a reflection test port as if it were a passive device.
- Keep a fixed frequency sampling path.Put the precision load on the main path and connect the counter through suitable directional sampling or a verified isolated splitter. Do not use a tee to put two 50 Ω receivers directly in parallel on the main path to view the waveform at the same time. Two ideal 50 Ω loads give 25 Ω, |Γ| = 1/3 and VSWR = 2. That arrangement can change the condition being tested. Where authorised for the oscillator, vary mismatch and phase within the manufacturer's permitted range and record the largest frequency offset and how it was reached.
- Check for a spurious reading.After the prescribed warm-up, measure with the precision load, the actual load and the precision load again, using the same sampling path. If temperature changes, check the difference between the two baseline readings. Record pin supply voltage, current and counter input power. Where necessary, vary the instrument level on the coupled branch without changing the actual load, to see whether the apparent offset is due to triggering.
When it is hard to improve matching, a pad can provide a diagnostic. For an ideal two-port matched 6 dB pad, the forward signal loses 6 dB and the reflection loses another 6 dB on the return. This reduces the load reflection at the oscillator by about 12 dB. The pad and cable also change its phase. Before interpreting a frequency change, verify that the instrument still meets its minimum input power, the receiver meets its threshold and the signal-to-noise ratio is sufficient. Real pads have their own return loss and power ratings. Qualification should use the reflection and frequency measured in the assembled configuration.
If there is a relay, RF switch or receiver that can power down, measure the states used in the actual equipment. The steady-state test should at least cover the receiver powered and settled, in standby and after switching. Capture a short phase jump during switching with a separate time-domain phase record. Counter gate averaging can leave only a small apparent shift. If two receivers present different loads in one state, establish whether the connectors and transmission lines differ in reflection or the input termination changes when power is removed. If the complete system is subject to hot plugging, the test method must specifically allow that condition.
Set the precision-load reading as the zero for load-induced shift. For example, if the baseline is 2.4 Hz above nominal and the customer load is 3.2 Hz above, the difference under identical temperature and supply conditions is +0.8 Hz or +8 ppb. Compare that value with the load-variation requirement. Retain the absolute equipment offset of +32 ppb separately for comparison with the initial frequency limit. Keeping both fields in the acceptance record shows which value the budget used when the counter or cable is changed later.
A supplier acceptance request can specify a 50 Ω baseline, actual receiver states, permitted frequency offset and minimum output power. Attach the complex reflection at the carrier along with temperature and supply records. Confirm that each second source covers the same load conditions. Checking the output interface, oscillator load pulling and test sampling chain separately makes it possible to choose between changing the receiver termination, adding buffer isolation and replacing the oscillator with one less sensitive to the actual load.