An active oscillator is specified at 20 MHz with a room-temperature offset limit of ±0.5 ppm. The measured frequency is 20,000,008 Hz. The reading is 8 Hz high and remains below the 10 Hz drawing limit. If the setup has an expanded uncertainty of 4 Hz, a direct comparison of the reading with the limit and a decision requiring the complete coverage interval to fit inside the specification give different answers. The inspection record needs explicit boundaries for release, retest and rejection.
What an incoming inspection record should show
These numbers illustrate a three-way rule the supplier and buyer can agree in advance. The drawing specifies a two-sided tolerance of ±0.5 ppm. Multiplying by 20,000,000 Hz and dividing by one million gives ±10 Hz, or 19,999,990 to 20,000,010 Hz. First correct the measured result for known systematic effects, then state its expanded uncertainty U. This example uses ±4 Hz. Release the device if the coverage interval lies wholly within tolerance, declare nonconformity if it lies wholly outside, and hold the device if it crosses the boundary. Agree in advance on the equality convention for an interval that touches the limit. In this example, contact from the inside allows release and contact from the outside remains a hold.
| Corrected reading | Coverage interval relative to nominal | Example disposition |
|---|---|---|
| 20,000,005 Hz | +1 to +9 Hz | Release |
| 20,000,008 Hz | +4 to +12 Hz | Hold for retest on the agreed setup |
| 20,000,011 Hz | +7 to +15 Hz | Hold under the same rule |
| 20,000,015 Hz | +11 to +19 Hz | Declare nonconformity |
Taking the guard band w = U gives an immediate release interval of ±6 Hz around nominal. Under the example equality convention, only an absolute offset greater than 14 Hz can be declared nonconforming on the initial setup. The intermediate region calls for retest. Apply the same logic to the lower limit so the test program checks both offset signs. JCGM 106, Section 8.3 describes simple acceptance and guarded acceptance, including the common case w = U. The three-way disposition in the table is an example contract procedure. If production needs a binary disposition, agree separately on how a hold will be resolved. Different decision rules carry different risks of false acceptance and false rejection. A frequency reading alone cannot establish the applicable rule.
Where the 4 Hz came from
Assume a simplified uncertainty budget for the room-temperature station. Convert the reference calibration certificate’s expanded uncertainty to standard uncertainty using its coverage factor, and include estimated drift since calibration. Use repeated remounting tests to separate fixture effects from reading repeatability. Then make small controlled changes in ambient temperature and supply voltage to estimate how closely the bench reproduces the specified test conditions. After grouping, the illustrative standard uncertainties are 1.3, 0.7, 1.1 and 0.7 Hz for the reference, reading repeatability, fixture remounting and condition reproducibility, respectively. If the components are approximately independent and have not been counted twice, combine them by root sum of squares. The combined standard uncertainty is about 1.97 Hz. A coverage factor of k = 2 gives about 3.94 Hz, rounded outwards to 4 Hz for this example.
When several stations share one timebase, averaging their readings cannot cancel its correlated error. Repeated readings of one oscillator can reduce some short-term statistical noise, but reference drift, an incorrect test supply voltage and noncompliant termination still need their own treatment. If the method permits correction of a known systematic effect, retain the corrected result and the uncertainty of the correction. The Keysight 53200A series data sheet lists random, systematic and timebase-related specifications for frequency measurement separately. Check the definitions and conditions behind those numbers before building the budget. The table entries cannot all be treated as independent one-standard-deviation values. Extra decimal places display smaller differences between readings.
Define the measurand before accounting for temperature. For initial offset at a specified room temperature, voltage and load, the relevant contribution is uncertainty in reproducing that test state. A contractual stability limit over the entire operating temperature range calls for a separate temperature test or qualification evidence covering that requirement. Putting the supplier's full-range stability specification into the room-temperature bench budget obscures both requirements. Likewise, a room-temperature pass says nothing about what happens after a cold start.
Where to send a unit on hold
In the purchase order or inspection procedure, specify the test point, load, reference and calibration interval. Define supply voltage, temperature and warm-up. Record the reading, correction, U, coverage factor, decision rule version and serial or reel lot number. Designate a retest station, the number of measurements and how they will be combined. If that station demonstrably reduces U to 1 Hz and measures the same unit at +8 Hz under the same conditions, the interval becomes +7 to +9 Hz and the example rule allows release. If the retest reads +10 Hz, the interval still crosses the boundary and the agreed borderline disposition applies. Measuring repeatedly until a lower number turns up undermines the agreed basis for the decision.
To make the decision software auditable, test at least three simulated units on each side of nominal, covering release, hold and nonconformity near both limits. Exercise the equality cases separately, including +6 and +14 Hz in this example. Use unrounded values internally. The display can use an agreed number of digits, but retain the original result for review. A borderline decision applies to the individual unit. The agreed lot sampling plan determines whether discovering that unit in an inspection sample puts the entire reel on hold.
If U on the initial station approaches the 10 Hz tolerance half-width, the guard bands consume nearly the entire release interval. Tightening the acceptance limits further will send many usable units to retest. First establish whether the reference, fixture repeatability and cycle time can support the required grade. When necessary, move the higher-accuracy decision to a more capable station and use documented uncertainty to verify the improvement. In volume shipments, having a disposition both buyer and supplier can reproduce does more to prevent borderline disputes than another displayed digit.