Measuring oscillator duty cycle? Check the passive probe compensation first

An active oscillator produces a flat-topped square wave at one station and a rounded shoulder at another. Both oscilloscopes report the correct frequency, but the second station rejects the duty cycle. Move the probe from the oscillator to the scope’s calibration square wave. If the calibration waveform shows the same rounded shoulder, there is already a lead to follow without touching the board. If the probe was recently moved to another channel, paired with another scope, or fitted with a new tip attachment, take a screenshot so the next operator can verify the compensation.

The compensation adjustment belongs to the measurement channel

A common 10× passive probe forms a divider with the scope input. The DC division ratio is mostly set by resistors. At higher frequencies, capacitance in the probe and scope input also affects the ratio. Compensation brings the low- and high-frequency division ratios into agreement. A square wave contains a range of frequencies, so a mismatch changes the displayed shape. With undercompensation, the edge may appear to approach the plateau slowly. With overcompensation, the signal may first overshoot and then settle back. Tektronix illustrates all three cases together in its probe primer and recommends rechecking after changing accessories. A 10× label gives the nominal attenuation. Compensation still has to suit the current probe and input combination.

For a periodic square wave, measure the interval between two successive edges of the same polarity at the same threshold. Even if the displayed edges shift, the reported period may stay unchanged. High time is measured from a rising crossing to the next falling crossing. When the two crossings move by different amounts, so does the reported high time. The threshold can move too. If the scope estimates the high and low plateaus automatically and sets the threshold to their midpoint, distortion can change both plateau estimates. Save the threshold definition and inspect the two crossings. That helps establish whether the change comes from the plateaus, the edges, or both.

Use the calibration output to leave a reproducible record

Check the inputs the probe manual allows. A normal 10× passive probe usually needs a compatible high-impedance input and the correct attenuation setting in the scope channel menu. A previous user may have left the channel in 50 Ω mode. That changes the divider conditions in a way the compensation adjustment cannot correct. Follow the probe and scope manuals to connect the tip to the calibration output and the ground to its nearby reference. Choose a range that shows several whole cycles with both plateaus visible. Look at the rising and falling edges. Adjust compensation or run the digital compensation procedure as instructed until the plateaus are flat. If the adjustment cannot get there within its allowed range, check specified input capacitance, probe tip, grounding accessory and channel settings. Try a known-good equivalent combination if necessary to isolate the fault.

Worked example with assumed values. A clock with a 100 ns period goes into a downstream circuit. The receiving inspection drawing specifies high time at a fixed voltage threshold. Assume a correctly compensated channel shows a rising crossing at 20 ns and a falling one at 70 ns. High time is 50 ns and duty cycle is 50%. Now assume the same signal appears to cross at 24 ns and 68 ns through a misadjusted measurement path, while successive rising crossings are still 100 ns apart. The displayed duty cycle is 44%. Substitute actual cursor readings to repeat the calculation. The direction of the error depends on the edge shapes, compensation and threshold. Six percentage points is specific to the example. Save separate screenshots of the calibration waveform and the oscillator. Record the channel, probe ID, attenuation and threshold. Capture the oscillator again under the same conditions after compensation.

Return to the board and check the acceptance conditions

The calibration square wave checks the probe and input pairing. The actual board adds probe capacitance, the test-point branch, ground lead inductance and the load specified for the part. Probe capacitance may change the real waveform of a small oscillator with little load margin. Connect the specified load and, if needed, use an appropriate low-capacitance probe or a buffered test point. Ground close to the output and record the position of the test point relative to the device. A flat calibration waveform is a starting point. Verify that the scope has the bandwidth and sample rate needed for the edge. Keep the compensation record with the probe and channel combination used to create it.

For automated measurements, check how the instrument defines reference levels and pairs edges. A component specification might use the waveform midpoint or a defined fraction of the supply voltage. To compare instruments, apply the specified threshold to the same exported waveform, recompute one period and its high time with cursors, and check against the automated report. If brief ringing crosses the same threshold several times, pass along the raw trace so operators can see which crossing each scope used. Retest borderline units at a verified station using the same retained sample. Keep the first result so earlier measurements can be reviewed for the same setup issue.

Include supply voltage, actual load, threshold and test point on the incoming inspection report. Identify which probe goes with which channel so the next shift can check again after changing equipment or accessories. Once the calibration trace is flat and both stations agree within expectations using the same definition, compare the result with the part-number limit. If the measurement setup checks out and distortion occurs on only one board, compare its output, loading and routing. Use the verified conditions to choose the station for the formal record.