An oscillator overshoot looks flat on top. How can you tell whether the scope clipped it?

To see a rising edge more clearly, an operator chooses a sensitive vertical range and moves the upper plateau to screen center. A single acquisition shows a flat-topped peak, with a stable automated maximum. Widen the acquisition range and check the raw record and instrument overload status. Once the analog front end or ADC exceeds its dynamic range, acquired samples may be limited to a maximum code. Bringing the trace into view may only reposition already acquired data. An overshoot measurement needs sufficient range at acquisition.

Separate hardware acquisition range from display zoom

The vertical range changes the voltage that reaches the converter. Some inputs also apply DC offset before acquisition. The complete waveform, including extrema and spikes, must fit the specified range and offset. Moving a trace afterward may only change its display position. Zooming a stored waveform is also a display operation. Neither recovers samples lost to clipping. Keysight’s Infiniium MXR/EXR programmer’s guide describes converter clip detection and notes that a channel range or offset change resets clip state. Have the script save each acquisition’s flags before changing settings.

A trace drawn beyond the screen edge does not establish clipping. An instrument can retain data outside the view that still fit the acquisition range. Read the model’s documentation and export the original acquisition without another display transform. If the peaks all have identical codes and the flat level moves when the hardware range changes, suspect a front-end limit. If two ranges agree in voltage and another qualified probe path shows the same flat top, investigate device behavior, circuit clamping or load. Check concurrent channel overload too. In a two-channel subtraction, a single clipped input can corrupt the result.

Repeat at two vertical ranges

Worked example with assumed values. Assume an effective input range of 0 to 2.0 V at a given offset. The actual normal high is 1.8 V and an occasional overshoot reaches 2.3 V. If all samples above the limit are returned as 2.0 V, the result shows only 0.2 V overshoot. On a range that captures 2.3 V and any negative ringing, suppose the high level remains 1.8 V. The overshoot is then 2.3 − 1.8 = 0.5 V. These numbers can be used to test inspection software. Choose a source whose peak and lower level meet instrument and probe ratings. Do not exceed those ratings to induce clipping.

Keep trigger, attenuation and test point the same. Start wide enough to see levels, peaks and the whole record, then narrow the range in steps. Reacquire on each setting, read the clipping flag and see whether automatic peak-to-peak voltage can be reproduced from the data. If an anomaly occurs only once, save its full original record. Comparing separate acquisitions requires a recurring event. A known analog test signal with a narrow step can show whether the station correctly reports the peak across ranges and expose unnoticed automatic changes.

When to replace the probe or measurement path

Clipping can occur in the probe. Active probes have offset, common-mode and differential limits. The probe may compress or saturate without an ADC clipping alarm. Check allowed input for the current attenuation and offset. Use suitable attenuation, another range or a specified direct connection if needed. Take account of bandwidth and loading so a higher-voltage setup does not substantially slow the edge. For a small differential swing on high common-mode voltage, verify instantaneous limits of each input relative to ground as well as the differential voltage.

If a receiving report uses output extrema to check the receiver’s ratings, ask for maximum, minimum, input range, probe status and raw waveforms for the same run. A clipped reading is evidence that the acquisition limit was reached. Retest on a wider range for the valid peak. If range is adequate and the flat top persists, investigate which circuit stage creates it. Export scripts can log the clip flag with each channel’s numerical result.

Shared channels or acquisition resolution may change range behavior. Fix the acquisition mode when multiple users test on one station, and check that changing voltage range did not also change bandwidth or sample rate. If bandwidth changed, select settings common to both ranges and repeat with the same verification pulse.

Set an initial range for each board. If output or supply changes with a revision, capture the full swing again and confirm that the more sensitive jitter and overshoot range still has headroom. Save both raw channels for subtraction reports. If any part clipped, reacquire after adjusting the range. A screenshot cannot recover data.