While measuring differential oscillator edges, the scope samples every 200 ps but reports a 30 ps difference. Software can estimate where a waveform crossed a threshold between its recorded samples. This is interpolation. A displayed change smaller than the sample interval does not establish equal accuracy. Edge shape, sample noise, timebase and delay in each channel affect confidence.
Estimate a crossing from raw samples
If the edge is nearly straight across a short interval, interpolate between the two voltages and their times. Suppose voltage rises 0.4 V between consecutive samples, with the threshold 0.1 V above the earlier sample. A linear estimate puts the crossing a quarter of the way into the interval. Some scopes use an interpolation model for band-limited signals. Tektronix describes different waveform assumptions for linear and sine-based interpolation. Display interpolation changes the drawn trace. Check the model’s algorithm for internal timing measurements separately. Measuring pixels on a screenshot cannot establish that algorithm.
The steeper the edge, the smaller the timing effect of a given voltage error. A rounded slope or ringing near threshold can change slope and make timing unstable. Channel gain differences also change the actual input voltage corresponding to a fixed numerical threshold. Deskew the channels and probes as recommended and check bandwidth before subtracting edge times. Some common timebase effects can cancel for simultaneous edges. Channel aperture skew, connection delays and measured signal noise can remain.
Worked example with assumed values. Samples at 0 and 200 ps are 0.4 V and 0.8 V on one edge. At a 0.5 V threshold, linear interpolation gives 50 ps. On a second edge, samples are 0.34 V and 0.74 V at the same times. The estimated crossing is 80 ps. The 30 ps difference follows even though it is below the sample interval. If random voltage noise near threshold is about 20 mV and slope is 2 mV/ps, local conversion gives about 10 ps of time uncertainty from that source. Each edge contributes uncertainty. Display decimals cannot establish the accuracy of the 30 ps result. Split one edge between two channels, repeat the arithmetic from exported voltages and sample times, and compare with what the scope reports for nominally simultaneous edges.
Check whether the estimate follows sample placement
For two channels fed by the same source, leave connections fixed and save repeated acquisitions. Slightly vary trigger position relative to the sampling clock, or use another approved sample rate, so the crossing lands at different fractions of one interval. A straight line interpolates independently of starting position. A curved waveform can vary with local sample positions. Denser actual samples near threshold can reveal whether automated estimates converge. For timing that must be simultaneous, verify real-time acquisition. Equivalent-time acquisition reconstructs multiple repetitions. A first edge or nonperiodic fault needs single-acquisition capability.
Use a known length difference in equivalent cable or a suitable calibrated delay device to add a known delay. Measure zero difference, then known difference, then the board. At both settings, keep probe and channel arrangement constant and document range, threshold, sample rate, bandwidth and wiring. Display software can smooth a few samples. Log how the calculation script used actual samples and save raw data and connection diagrams so another lab can apply the same definition.
Decide whether 30 ps belongs on an acceptance report
A screening method meant to catch discrepancies of several hundred picoseconds may be adequate with the current noise and repeatability. If a contractual limit approaches noise and residual skew, improve the setup or use another method. Check the spread after repeatedly reconnecting a retained sample and whether a channel swap reverses sign as expected. Compare the sample with the verified connection. Save display resolution, repeatability, instrument timing specification and calibration evidence separately.
Triggering all records at the same position can hide interpolation sensitivity. Shift the phase between sync and signal so the edge traverses a full sample period, using the same edge and termination. Compare measurements at different relative positions. If the reported difference repeatedly increases at one position, log it as another error source.
Define how to read a single-record timing difference and how many repeats to use in station qualification. For rare maxima, keep every raw record. An averaged waveform reduces noise and may wash out real changes on a small number of edges. Use an estimate that is repeatable under the agreed definition.