Two labs measure the same oscillator. Carrier and discrete spurs largely agree, but the broadband floor on one plot is 2–3 dB lower. Check how each spectrum analyzer averages random signals at each frequency. Averaging linear power before converting to decibels can give a different result from converting individual samples to logarithms first. A steady tone may look almost the same either way. Random noise fluctuates, so the order can change the floor on screen. For a low-noise oscillator near its acceptance limit, a few decibels can change the disposition.
See how the spectrum trace was calculated
At a displayed frequency, an analyzer can acquire several internal samples and form a display value with a selected detector. It can also average traces across sweeps. The marker may report power directly or convert to a per-hertz value using the equivalent noise bandwidth. These settings are often grouped in instrument manuals. Record whether the test uses power averaging, log averaging or ordinary peak detection. The Keysight FieldFox spectrum analyzer help gives examples of how noise traces vary by mode. Under its stated mode and noise conditions, power-averaged random noise can read about 2.5 dB above log averaging. Identify the setting behind each result.
Close-in continuous noise, a discrete line and analyzer noise can appear together. If a marker is near a line, changing resolution bandwidth can alter its visible shape and floor. First identify whether the quantity is discrete line power or bandwidth-normalized noise density. To compare the floor, choose a band free of spurs and check whether the marker reports relative or absolute power. Reference level, attenuation, preamp and differences in carrier level also affect system noise. Align averaging first, then control these settings.
Worked example with assumed values. Suppose two independent samples at one frequency give linear power of 1 and 9 units. Averaging power gives 5 units, or about 6.99 dB relative to 1 unit. Converting the individual readings gives 0 and about 9.54 dB. Averaging those values gives about 4.77 dB, a difference of about 2.22 dB. The two numbers illustrate calculation order. They do not estimate the average bias for an actual random noise process. To recalculate from exported data, check what quantity the file holds, whether noise marker correction was applied and whether the instrument already averaged internal samples. Do not label a further average of a log-detected display trace as linear power.
Retest with one variable under control
Keep the same oscillator, offset frequency, input power and termination. On one analyzer, fix resolution bandwidth, video bandwidth and sweep settings, then change only averaging mode. Give both traces the same acquisition time. Export the complete traces and settings if possible. Check the white floor near a stable carrier or use a verified noise source to avoid putting an oscillator spur or drift inside the marker window. Record sample count and statistic, and mark the measurement band on the plot as needed. Close to the carrier, a broad window can combine very different noise levels across offset frequencies. Compare one offset first, then consider an average across a band.
For another analyzer, use its manual to select a mode that reports the same physical quantity. Some noise markers already correct for equivalent noise bandwidth and detector effects. Applying an empirical correction again adds error. For a per-hertz result, modest changes in resolution bandwidth should produce broadly similar readings in a flat noise band once corrected appropriately. If a line or carrier skirt is nearby, check the marker’s continuous background content. For tests close to the analyzer floor, include a baseline for the measurement chain and report the margin at that offset.
Clear accumulated averaging whenever a setting changes, so traces from the old and new configurations do not mix. Other stations can load the same setup and restart acquisition, checking that the software did not revert to the default detector. When a specification limits noise at a particular offset, agree on the marker definition beforehand. A retained sample below the limit or a suitable source can provide a daily baseline to investigate a shift across all production units. Keep the reference, acquisition duration and version with the trace. Give the supplier before-and-after traces at the same input power and bandwidth, separating evidence about averaging from variation across samples. Record the criterion for thermal stabilization when using a retained sample.
Acceptance can list continuous noise, maximum spur and carrier power separately with instrument settings and units. When two reports differ by a few decibels, ask for a retest of the same retained unit using one agreed averaging method. If the difference follows the mode, save the lot results with the specified mode. If it remains after matching the settings, return to instrument floor, connections and reference status before deciding whether the oscillator actually changed.