Why three 10 MHz OCXO comparisons can yield a negative Allan variance

Three 10 MHz oven-controlled crystal oscillators have been compared in pairs. The report contains three Allan deviation curves. The technician applies the three-cornered hat and finds a negative individual variance at one averaging time. Purchasing wanted to identify the best unit, but the spreadsheet cannot take the square root. Preserve the negative value and inspect the raw data. First check whether the three comparisons cover the same operating interval, then look for two sources drifting together.

What the three-cornered hat assumes

Gray and Allan’s original paper estimates the stability of three oscillators from simultaneous comparisons. At each averaging time τ, let VAB, VAC and VBC denote the Allan variances of the pairwise differences. If the three sources’ frequency fluctuations are approximately uncorrelated on this scale and the measurement system contribution is negligible, VAB is approximately VA + VB, with equivalent relations for the other two pairs. Solving gives VA = (VAB + VAC − VBC)/2. Start with variances. If the report supplies Allan deviations, square each first and take square roots only after obtaining nonnegative variance estimates.

For example, with variance in units of 10−22, three values at τ = 10 s are 10, 17 and 25. The estimated variances of A, B and C are 1, 9 and 16 in those units and the deviations are 1, 3 and 4 × 10−11. If another record gives pairwise variances of 4, 9 and 16, the same calculation gives A as −1.5 × 10−22. Taking the absolute value and square root turns a diagnostic into a small and apparently excellent device result. Replacing it with zero also biases the comparison. Mark the individual contribution as unresolved and keep the three pairwise curves and sample counts for review.

Check the timeline before tracing shared influences

Do all three differences come from the same interval?
Put all three raw phase records on a common timeline and use the same uninterrupted interval and estimator. Measuring A versus B on Monday, A versus C on Tuesday and B versus C on Friday can mix equipment cycles, environment and aging states. Using one instrument for all three does not make them simultaneous. Use synchronous multichannel phase records, or derive the third difference from two synchronous measurements against a common reference while accounting for channel noise and the reference’s role.
What does a closure check establish?
If the three pairwise differences were recorded independently, check xAB + xBC − xAC with consistent phase signs. It should be near zero at the same epoch, subject to independently calibrated path offsets and measurement noise. A whole-cycle step calls for checking timestamps, signs, cabling and cycle slips. If the third difference was calculated from the first two, closure holds by construction and cannot verify acquisition accuracy. Keep common-source floor data and an independent channel comparison.
Do the oscillators respond to a shared event?
A fan switching in one rack, a shared supply sag or two units steered by the same receiver’s time pulse can create correlated changes. With CAB defined as Allan covariance, the pairwise result includes −2CAB. If A and B speed up together relative to their individual means, their difference shrinks. More subtly, if all three share the same frequency disturbance, it can disappear from every pairwise record. Positive decomposed estimates still cannot see that common change.

Premoli and Tavella’s study uses a covariance model to address negative values from the original method. If using such a method, state the extra constraints, record duration and environment. Three differences alone cannot identify arbitrary common fluctuations. On the bench, first isolate mechanical and supply coupling, verify lock states and compare with an independent reference. Then decide whether the estimates support a device ranking. If two oscillators are substantially noisier than the device, subtracting large estimates magnifies uncertainty. A longer run still needs enough effective difference terms.

A numerical example shows how common noise changes the separation. Suppose the independent Allan variances of A, B and C are 1, 9 and 16. Let B and C share an additional disturbance, independent of those fluctuations, with variance 4. The pairwise variances become 14, 21 and 25, because the shared term cancels in B−C. The hat estimates 5 for A, even though A never received the disturbance. When A instead correlates with its references, its own variance can be underestimated. Trace which oscillator and which measurement path each shared event enters before judging the direction of error. Synchronous timestamps help align observations and do not remove correlations in the oscillators themselves.

Before acceptance, recalculate on separate time segments with the same mounting. Report daytime steady state, overnight steady state and air-conditioning transitions, for example, then test whether conclusions repeat over intervals of a fixed duration. Keep the original traces and settings when changing the analysis so an unfavourable interval does not disappear silently. If operation necessarily includes those transitions, evaluate them separately as service conditions. Errors in the three pairwise variance estimates can themselves be correlated because they share clocks and sampling times. Do not add their error bars as independent quantities. Use a statistical method that accounts for shared data or collect independent repeat runs.

Leave room to report an unresolved grade

Record all three pairwise variances, the estimates, valid duration, supply and temperature, τ and statistical method. At long τ, the effective degrees of freedom can be much smaller than the total sample count. The first few stable values cannot establish long-term repeatability. If units can be separated at 1 s and 10 s but the 1,000 s comparison is inconclusive with two references responding to room temperature, report each result at its applicable interval. A separately verified lower-noise reference can retest the interval that matters for purchase.

If the contract specifies both stability at a chosen averaging time and frequency accuracy, report both. The differenced variance describes fluctuations. All three units could miss nominal by several tens of millihertz while three-cornered hat estimates still show small random variation. Accuracy needs a separate calibration point from a traceable frequency reference. Keeping the two separately reproducible results shows which claims are supported after changing the reference or adding an inspection station.